Reprocessing endoscopes: Ch. 9 Exam All Possible Questions and Answers with complete solution
the most critical step in the manual reprocessing of flexible endoscopes is - Answer-Cleaning reprocessing is considered "delayed" if a scope is not cleaned within - Answer-60 Mins The purpose of precleaning at the point of use is to - Answer-prevent the buildup of bioburden and the formation of biofilm precleaning at the point of use involves which steps - Answer-Flushing of the channels containers or bags used to transport contaminated scopes should be labeled - Answer-As biohazardous before a flexable scope is manually cleaned, it should be - Answer-leak tested coiling a scope too tightly may cause - Answer-Damage to the fiber optic rods Which is NOT a method of leak testing - Answer-Suction A detergent that is too concentrated can - Answer-cause inadequate rinsing After each use reusable cleaning brushes must be - Answer-disinfected endoscopes should be inspected for cleanliness - Answer-using a lighted magnifier