Fluoroscopy exam Questions and Answers- Updated 2024 Total filtration permanently required in the useful fluoroscopic beam should not be less than - Answer-2.5mm Aluminum If X-ray source is below the table, useful beam exposure must be measured at what distance above the table top - Answer-30 cm Table top exposure rates for units without auto exposure control must be measured every - Answer-3 years Dark adaptation for the eye takes how long - Answer-30 minutes Target-to-tabletop distance must be greater than and should be greater than... - Answer-12 and 18 inches Integration time for brain to recognize an object - Answer-0.2 seconds Most scatter occurs at which direction from the primary beam - Answer-135 degrees mA setting for spot filming where a spot film camera is not used - Answer-150 mA Greatest dose to the patient comes from spot filming cameras using this film size - Answer- 105 mm EMILLECT 2024/2025 ACADEMIC YEAR ©2024 EMILLECT. ALL RIGHTS RESERVED FIRST PUBLISH OCTOBER, 2024 Page 2/16 SI unit analogous to the rad is - Answer-Gray _________ is a variation in density resulting from the random spatial distribution of quanta. - Answer-Quantum sink SI unit analogous to the rem - Answer-Sievert (unit) Exposure reduction of .25 mm lead - Answer-97% Scattered radiation received is proportional to - Answer-Dose received by the patient Most effective medium for static viewing in terms of reducing patient dose - Answer-Video disc recorder Contrast of the image intensifying tube is measured by - Answer-conversion factor maximum magnification for cinefluorography is done via which method - Answer-total over framing Ratio of photons that leave the output phosphor to the number of Xray photons that strike the input phosphor - Answer-Flux gain Half value layer determines which characteristic of the Xray - Answer-Xray quality Ability to distinguish separate objects in line pairs per mm - Answer-Resolving power The portion of the image intensifier responsible for magnification - Answer-electrostatic lenses resolution of a 9 inch image intensifier tube - Answer-3.5 line pairs per millimeter EMILLECT 2024/2025 ACADEMIC YEAR ©2024 EMILLECT. ALL RIGHTS RESERVED FIRST PUBLISH OCTOBER, 2024 Page 3/16 light photons intercepting when electrons strike a fluorescent screen forms a... - Answer- Real image Fluoro with a grid in place affects exposure rates how - Answer-1.5-2 x higher Greatest radiosensitivity of fetus occurs when - Answer-2-8 weeks gestation The quantity of the beam is controlled by which unit of measurement - Answer- miliamperage Which fluoroscopic method gives patient dose reduction of up to to 95%? - Answer-video disc recorder 100erg/gram is equivalent to? - Answer-1 rad Fluoro units with AECs or CINE cameras must have physicist calibration how often - Answer- every year Which imaging modality has the best resolution - Answer-Radiograph Compared to the input phosphor, the image output phosphor is... - Answer-Smaller and brighter As spot film size increases, what happens to the patient dose? - Answer-Patient dose increases Unit used to specify the quality of an X-ray beam - Answer-HVL

Meer zien Lees minder
Instelling
Fluoroscopy
Vak
Fluoroscopy










Oeps! We kunnen je document nu niet laden. Probeer het nog eens of neem contact op met support.

Geschreven voor

Instelling
Fluoroscopy
Vak
Fluoroscopy

Documentinformatie

Geüpload op
16 oktober 2024
Aantal pagina's
16
Geschreven in
2024/2025
Type
Tentamen (uitwerkingen)
Bevat
Vragen en antwoorden

Onderwerpen

Voorbeeld van de inhoud

EMILLECT 2024/2025 ACADEMIC YEAR ©2024 EMILLECT. ALL RIGHTS RESERVED FIRST PUBLISH OCTOBER, 2024




Fluoroscopy exam Questions and
Answers- Updated 2024

Total filtration permanently required in the useful fluoroscopic beam should not be less than -

Answer✔✔-2.5mm Aluminum


If X-ray source is below the table, useful beam exposure must be measured at what distance

above the table top - Answer✔✔-30 cm


Table top exposure rates for units without auto exposure control must be measured every -

Answer✔✔-3 years


Dark adaptation for the eye takes how long - Answer✔✔-30 minutes


Target-to-tabletop distance must be greater than and should be greater than... - Answer✔✔-12

and 18 inches


Integration time for brain to recognize an object - Answer✔✔-0.2 seconds


Most scatter occurs at which direction from the primary beam - Answer✔✔-135 degrees


mA setting for spot filming where a spot film camera is not used - Answer✔✔-150 mA


Greatest dose to the patient comes from spot filming cameras using this film size - Answer✔✔-

105 mm


Page 1/16

,EMILLECT 2024/2025 ACADEMIC YEAR ©2024 EMILLECT. ALL RIGHTS RESERVED FIRST PUBLISH OCTOBER, 2024


SI unit analogous to the rad is - Answer✔✔-Gray


_________ is a variation in density resulting from the random spatial distribution of quanta. -

Answer✔✔-Quantum sink


SI unit analogous to the rem - Answer✔✔-Sievert (unit)


Exposure reduction of .25 mm lead - Answer✔✔-97%


Scattered radiation received is proportional to - Answer✔✔-Dose received by the patient


Most effective medium for static viewing in terms of reducing patient dose - Answer✔✔-Video

disc recorder


Contrast of the image intensifying tube is measured by - Answer✔✔-conversion factor


maximum magnification for cinefluorography is done via which method - Answer✔✔-total over

framing


Ratio of photons that leave the output phosphor to the number of Xray photons that strike the

input phosphor - Answer✔✔-Flux gain


Half value layer determines which characteristic of the Xray - Answer✔✔-Xray quality


Ability to distinguish separate objects in line pairs per mm - Answer✔✔-Resolving power


The portion of the image intensifier responsible for magnification - Answer✔✔-electrostatic

lenses


resolution of a 9 inch image intensifier tube - Answer✔✔-3.5 line pairs per millimeter

Page 2/16

, EMILLECT 2024/2025 ACADEMIC YEAR ©2024 EMILLECT. ALL RIGHTS RESERVED FIRST PUBLISH OCTOBER, 2024


light photons intercepting when electrons strike a fluorescent screen forms a... - Answer✔✔-

Real image


Fluoro with a grid in place affects exposure rates how - Answer✔✔-1.5-2 x higher


Greatest radiosensitivity of fetus occurs when - Answer✔✔-2-8 weeks gestation


The quantity of the beam is controlled by which unit of measurement - Answer✔✔-

miliamperage


Which fluoroscopic method gives patient dose reduction of up to to 95%? - Answer✔✔-video

disc recorder


100erg/gram is equivalent to? - Answer✔✔-1 rad


Fluoro units with AECs or CINE cameras must have physicist calibration how often - Answer✔✔-

every year


Which imaging modality has the best resolution - Answer✔✔-Radiograph


Compared to the input phosphor, the image output phosphor is... - Answer✔✔-Smaller and

brighter


As spot film size increases, what happens to the patient dose? - Answer✔✔-Patient dose

increases


Unit used to specify the quality of an X-ray beam - Answer✔✔-HVL


cycles per time unit in a TV camera is known as... - Answer✔✔-bandwidth

Page 3/16
€9,18
Krijg toegang tot het volledige document:

100% tevredenheidsgarantie
Direct beschikbaar na je betaling
Lees online óf als PDF
Geen vaste maandelijkse kosten


Ook beschikbaar in voordeelbundel

Maak kennis met de verkoper

Seller avatar
De reputatie van een verkoper is gebaseerd op het aantal documenten dat iemand tegen betaling verkocht heeft en de beoordelingen die voor die items ontvangen zijn. Er zijn drie niveau’s te onderscheiden: brons, zilver en goud. Hoe beter de reputatie, hoe meer de kwaliteit van zijn of haar werk te vertrouwen is.
Emillect West Virginia University
Volgen Je moet ingelogd zijn om studenten of vakken te kunnen volgen
Verkocht
28
Lid sinds
1 jaar
Aantal volgers
2
Documenten
3004
Laatst verkocht
1 maand geleden
GOLDEN ORCHIDS STORE.

On this page, you find all documents, package deals, and flashcards offered by seller Emillect.

3,0

4 beoordelingen

5
1
4
0
3
2
2
0
1
1

Recent door jou bekeken

Waarom studenten kiezen voor Stuvia

Gemaakt door medestudenten, geverifieerd door reviews

Kwaliteit die je kunt vertrouwen: geschreven door studenten die slaagden en beoordeeld door anderen die dit document gebruikten.

Niet tevreden? Kies een ander document

Geen zorgen! Je kunt voor hetzelfde geld direct een ander document kiezen dat beter past bij wat je zoekt.

Betaal zoals je wilt, start meteen met leren

Geen abonnement, geen verplichtingen. Betaal zoals je gewend bent via Bancontact, iDeal of creditcard en download je PDF-document meteen.

Student with book image

“Gekocht, gedownload en geslaagd. Zo eenvoudig kan het zijn.”

Alisha Student

Veelgestelde vragen