______ is the only one that can be used directly for energy or
storage Correct Answers glucose
+/- 1SD Correct Answers 68.2 %
+/- 2SD Correct Answers 95.5 %
+/- 3SD Correct Answers 99.7%
1(2s) Correct Answers outside of +/- 2SD
warning rule
1(3s) Correct Answers outside of +/- 3SD
reject: random error
2 hour post prandial test Correct Answers -Patient is given a 75
gram dose of glucose.
-Two hours later, blood is drawn.
-If the level is > 200 mg/dL, and is confirmed on another day,
the patient is diagnosed with Diabetes Mellitus.
-less than 140 is considered normal
2 step glucose tolerance Correct Answers step 1
-50 g Glucose Load is given
-If > 140 mg/dL after one hour then proceed to Step 2
step 2
-100 g Glucose Load
-3 hour OGTT
,-Gestational Diabetes if: Fasting: >92 mg/dL , 1 Hour: > 180
mg/dL , 2 Hour: > 153 mg/dL , 3 Hour: > 140 mg/dL
2(2s) Correct Answers outside of +/- 2SD two days in a row
reject: systematic error
4(1s) Correct Answers four consecutive controls exceed the true
mean by +/- 1SD
reject: systematic
A shift is a change in ____, therefore it is a ______ error Correct
Answers accuracy, systematic
A trend is a change in ____ and therefore a ______ error Correct
Answers accuracy, systematic error
Absorbance calibration/photometric accuracy Correct Answers
Preformed daily
Purpose: ensures the absorbance reading is accurate and linear
To ensure a lack of stray light
Neutral density filter (340 nm , 500 nm) and cobalt ammonium
sulfate solutions
alcohol has a negative effect on glycogen storage Correct
Answers glycolysis
alimentary hypoglycemia Correct Answers -patients who have
had GI surgery
-accelerated absorption of glucose-yields an exaggerated insulin
response
, Alpha stereoisomer Correct Answers Carbonyl group (CO) is
on the right (look at top)
amylopectin Correct Answers alpha 1,6 linkages every 25 units
(branches)
amylose Correct Answers only alpha 1,4 linkage (spiral shape)
An outlier is a change in _____, and therefore a ____ error
Correct Answers precision, random
Analytical variation Correct Answers -Calibration,
documentation of protocol and procedure, monitoring equipment
and materials, preparation and storage of reagents, lot to lot
variation
-daily and monthly preventative maintenance
applications of nephelometry Correct Answers Quantitative
determination of various proteins and other antigens in blood,
serum, urine, or CSF
Determination of cell size
Amount of amnio acids, vitamins, and Abs
Determination of molecular weight
Bandpass are influenced by Correct Answers Type of
monochromator
Exit slit width
Distance between the monochromator and exit slit (greater the
distance the more narrow the bandpass)
Beers law formula Correct Answers A=abc