Questions and CORRECT Answers
Uses Cs-137 for calibration
Quality control and calibration for sodium iodine (NaI) Calibration = daily
scintillation detector Uniformity = quarterly
Linearity = annually
Resolution = at installation or repair
they help identify the energy of the gamma rays emitted
by radioactive sources
photopeak
Full width half maximum
Gamma ray spectra
Gamma ray spectra provide a detailed view of the radi-
ation detected by scintillation detectors and are crucial
for identifying radionuclides, quantifying activity, and per-
forming quality control in nuclear medicine
represents the energy of gamma photons that are com-
Photopeak pletely absorbed by the detector, usually after undergoing
interactions such as the photoelectric effect
a measure used to quantify the width of a peak in a
spectrum, important when measuring energy resolution
Full width half maximum (FWHM) FWHM tells us how "sharp" or "broad" a peak is and
provides insight into the detector's ability to distinguish
between different energies
technique used in gamma spectroscopy to analyze the en-
ergies of gamma rays detected by scintillation detectors
Pulse height analysis
detailed pulse height spectrum that can be used for iden-
, tifying radionuclides, quantifying activity, and performing
quality control
Formula energy resolution
Formula sensitivity
Formula Chi-squared statistics
Gm meter or portable ionization chamber
Functions by reading ionization of radiation
Survey meter operations
Expressed in mR/hr
Calibration = installation or repair
Battery check = daily
Gas filled, suflcient sensitivity for contamination
Survey meter components
operate in rate mode
Calibration = before first use, at repair
Survey meter QC
Battery check = daily
Daily
Dose Calibrator constancy
done using long lived source, done on all radiopharm
settings
annually
Dose calibrator accuracy
two sources used (typically Cs137 and Co57)
Dose calibrator linearity