100% de satisfacción garantizada Inmediatamente disponible después del pago Tanto en línea como en PDF No estas atado a nada 4.2 TrustPilot
logo-home
Resumen

Summary different methods/sources from Concepts of Protein technology and applications

Puntuación
-
Vendido
-
Páginas
14
Subido en
27-08-2025
Escrito en
2024/2025

This document is a summary in the format of a table where I explain all the different aspects that are present in the course Concepts of Protein technology and applications. It contains only the part of Prof. Van Ostade. Subjects present in this document are: ionisation sources, mass analysers, improvement of the resolution of TOF, detectors, 4 major MS/MS modes of operation, Hybrid instruments, protein identification, quantification methods on the basis of isotopes and label-free quantification methods. All these subjects contain an explanation of their principle, scheme, advantages and disadvantages. Result: 15/20.

Mostrar más Leer menos
Institución
Grado













Ups! No podemos cargar tu documento ahora. Inténtalo de nuevo o contacta con soporte.

Escuela, estudio y materia

Institución
Estudio
Grado

Información del documento

Subido en
27 de agosto de 2025
Número de páginas
14
Escrito en
2024/2025
Tipo
Resumen

Temas

Vista previa del contenido

IONIZATION SOURCES
Principle Scheme Advantages Disadvantages
Electrospray Ionisation • Apply strong electric field. • Practical mass limit • Low salt tolerance.
(ESI) • Redox chemistry at interface ~70,000 Da. • Di[iculty cleaning
= combined actoion of between solution and electrode • Good sensitivity. overly contaminated
repulsion and à charging of solution + • Softest ionization. instrument.
evaporation of the formation of Taylor cone. • Easily adaptable to • Low tolerance for
solvent • Repulsion of the ions à microbore LC. mixtures.
nebulisation (small droplets are • No matrix interference. • Multiple charging.
formed and there is a fast flow • Easy adaptability to • Quantitation requires
outwards. triple quadrupole internal standard.
• Counter current of (heated) gas, analysis.
• ESI source perpendicular to the mass
perpendicular to the spray à • Multiple charging.
analyser
evaporation of the solvent • Better mass accuracy.
• Multiple charging allows for calculation of
within the droplets. • Determine peptide
the MW
• Density of the surface charges modifications.
o z1*(P2-P1) = P2 – 1
increases à electrostatic • Samples can be
o M = (P1*z1) – z1
repulsion becomes higher than analysed directly.
surface tension à droplets • MS/MS capabilities.
explode into smaller droplets.
• These undergo the same
process, resulting in gradual
formation of smaller droplets.
• Happens under atmospheric
pressure.
MALDI (Matrix Assisted • Analyte is crystallized on a • Practical mass limit • Low resolution.
Laser Desorption stainless-steel plate, together with ~300,000 Da. • Matrix background.
Ionisation) an excess of matrix. • Highly sensitive. • MS/MS capability is
• Matrix is able to absorb a specific
• Soft ionization. minimal.
wavelength and is usually a weak
organic acid (proton donor) with a • Suitable to analyse • Photodegradation.
phenolic group. It can spread out complex mixtures. • Not possible to study
easily on the support. • Samples added directly noncovalent
• Interaction of laser light with the to appropriate matrix. interactions.
sample results in ionization of • Requires internal
matrix and analyte molecules. standard.
• A high potential electric field leads • Poor for determine
the ions to the mass analyser.
peptide modifications.
• Low tolerance for
various experiments.

, MASS ANALYSERS
Principle Scheme Advantages Disadvantages
Magnetic Sector • Ions are accelerated in an electric field
Analysers and pass two analysers.
• Electrostatic analyser (or ESA) functions
as a ‘kinetic energy filter’ and allows only
ions to pass with a specific kinetic
energy.
• Ions with diGerent masses and identical
speed arrive at the magnetic sector and
form a bended trajectory (deflection)
with a radius that is dependent of the
m/z value and the speed of the ion.
• Since the last one is the same for all the
selected ions, this means that only ions
of a specific m/z value form a bend that
allows them to hit the detector (m/z
filter). Adapting the magnetic field,
another set of ions will reach the
detector.
• Hence by gradually changing the
magnetic field from one extreme to the
other, ions are scanned from high MW to
low MW or vice versa.
Quadrupole • Four parallel rods. • Mass limit ±70.000 Da. • Scanning limits the time
• Combination of AC (rf region) and • High mass accuracy. to analyse ions within a
DC voltage is applied. On one pair of • Good MS/MS chromatographic peak
rods: negative voltage, on the other possibilities. (peak saturation).
pair: positive voltage. • Compatible with ESI
• Ions undergo oscillating movement sources.
when passing through the rods.
• Each pair makes up a filter: high
pass and low pass mass filters.
• Combination of these filters à
small window: only ions with a
specific m/z pass the quadrupole,
hence at a certain U, V and ω only
ions with a specific m/z value have
stabile oscillations while other ions
bend and are thrown out of the
quadrupole.

, • Scanning: this process repeats with
gradual changing U and V values
such that over time, di[erent ions
are detected.
• One scan = one duty cycle.
3D-iontrap • All ions enter the trap. • Di[erent analyses on • Space charge e[ects in
• Energetic ‘sink’ caused by RF field the same ion can be ion traps.
and helium. performed in the same
• Ions oscillate and accumulate into analyser.
the centre of the trap. • High sensitivity.
• Primary rf field is set to the ring • Possibility to study
electrode. Additional rf (or AC) is set • Adaptation of the quadrupole à 3- PTMs.
to the ‘end caps’. dimensional quadrupole built with 3
• Primary and additional rf are electrodes
gradually and simultaneously o 1 hyperbolic ring electrode
increased, with each ion (specific o 2 ‘end-cap’ electrodes which contain
m/z) resonating more and finally holes where the ions can pass.
thrown out for detection.
2D-iontrap • Based on 4 rods of a quadrupole, • More accurate.
with a positive electrode (lens) or • Volume much larger
two outside segments at each end. (more ions trapped).
Hence, radial trapping (x-axis) is • Better trapping
performed by the quadrupole, axial e[iciency (50% vs. 5%
trapping (z-axis) by the electric field with 3D-IT).
on the lenses.
• Ions move between the two outside
lens electrodes or segments (z-axis),
while they oscillate in x- and y-
direction. Ions can leave the trap in
axial or radial direction.

,FT-ICR (Fourier • Ions are injected into a small cube- • Good resolution and
Transform Ion shaped space, positioned into a very accuracy frequency
Cyclotron strong alternating fixed magnetic measurement
Resonance) field, applied on the z-axis. • MSn is possible
• Ions are introduced via the z-axis by
a trap voltage (1V) applied on the
back and front plate.
• All ions that are present in the
cyclotron are then excited (RF) after
which a fast (1 µs) scan in the
magnetic field over a broad
frequency range is applied.
• The ions rotate in the xy plane
around the z-axis so that they move
back and forward.
• Detector plates on both sides of the
cube measure the frequency by
which ions move through the tube
• Fourier transformer can convert this
complex signal into the individual
frequencies and intensities,
analogous to sound.
Orbitrap • Ions with high velocity (kinetic • Very high resolution
energy, 1600 eV) are tangentially (60.000-150.000
injected. FWHM), dependent on
• Inner electrode is adjusted to 3200V detection time.
à charged ions under influence of a • Not so expensive
centrifugal and an electrostatic • Accuracy: till ± 1 ppm.
force à ions circulate around the
inner electrode while moving along
the z-axis, also back and forth along
axis of central electrode.
• The frequency and the angular
velocity of these movements are
inverse proportional with m/z.
• Detection plates are present in the
outside electrode à similar to FT-
ICR.

,TOF (Time-Of- • Ions are accelerated by an electric Compared to quadrupoles: • Resolution.
Flight) field, set between an electrode and • No time-consuming • Accuracy.
an extraction grid. scan because of
• Ions are then introduced into a field- parallel detection of Improvements (see table
free vacuum zone. many di[erent ions underneath):
• The ion moves straight and with (µseconds), hence, • Delayed pulse
constant velocity to the detector. high throughput extraction.
• Smaller and multiple charged ions possibilities. • Reflectors.
move faster compared to heavier • Better suited for • Orthogonal
and monocharged ions. analysis of complex acceleration: ESI-TOF or
mixtures. QTOF.
• Sensitive.
• Mass limit: 300,000 Da.
• High salt tolerance.

, IMPROVEMENT OF THE RESOLUTION OF TOF
Problem Solution
Inbalanced dispersion of kinetic energy 1) Delayed pulse extraction
over ions with identical m/z • A voltage gradient is set with the highest V closest to the source (at the beginning)
• This happens during the MALDI desorption à that accelerates the fast ions less than the slow ions (which are closest to the
process. source)
o Ions expand in the field-free zone • Lower energy ions acquire more energy and catch up with the faster ones at the
and this expansion is dependent of detector.
their kinetic energy.
o For ions with identical m/z values, 2) Reflecors
the ions with higher kinetic energy • It increases the distance that the ion needs to travel.
move faster towards the detector • Consists of a gradient electric field
than ions with lower kinetic energy • Ions move inside this field according to their energy.
o Higher energy ions move faster thus deeper into the field and have to
travel a longer distance as compared to lower energy ions.
• At the end, times are focussed

3) Orthogonal acceleration: ESI-TOF or QTOF
• TOF tubes can be adapted for ESI ionization by placing them orthogonally to the ESI
source.
o ESI creates ions à focused in an ion beam that fills up the first part of an
acceleration chamber
§ This part is localized between a plate and a grid G1.
§ Initially no V is applied à ions can move freely away from the source.
o A voltage is applied on the plate causing the ions to move into a direction,
perpendicular to the ion beam.
o Ions are further accelerated towards the second grid G2 by means of a VTOF
voltage and separation can take place.
o While moving inside the TOF tube, the acceleration chamber can again filled
with new ions.
• Main advantages
o Online connection with ESI source or quadrupole
o Di[erent initial velocities are minimized after changing direction, resulting in better resolution and accuracy
$7.67
Accede al documento completo:

100% de satisfacción garantizada
Inmediatamente disponible después del pago
Tanto en línea como en PDF
No estas atado a nada

Conoce al vendedor
Seller avatar
WillemsenAmber
4.0
(1)

Conoce al vendedor

Seller avatar
WillemsenAmber Universiteit Antwerpen
Seguir Necesitas iniciar sesión para seguir a otros usuarios o asignaturas
Vendido
9
Miembro desde
1 año
Número de seguidores
0
Documentos
42
Última venta
1 día hace

4.0

1 reseñas

5
0
4
1
3
0
2
0
1
0

Recientemente visto por ti

Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes