What would increase the resolution of EEG spikes and improve
definition of the waveforms?
a. applying 50/60 Hz filter
b. using a 100 uV input voltage
c. increasing sensitivity
d. increasing sampling rate - Correct Answer - increasing sampling rate
Advanced digital acquisition systems samples continuous signals at a
fixed rate, called a sampling rate. The Nyquist theorem states that a
digital copy of a signal can be completely reconstructed from samples
made at a sampling rate faster than twice the bandwidth of the highest
frequency of the original data. Sampling at a higher rate facilitates more
accurate reconstruction of the signal. EEG spikes, such as those seen in
epileptiform activity, require a high sampling rate to reconstruct the
signal, enhance resolution and improve the definition of the waveform.
The best candidate for oral appliance therapy would have?
a. Moderate RBD
b. End-stage COPD
c. Severe snoring
d. Mild OSA - Correct Answer - mild OSA
pg. 1
,2|P a g e
Which of these factors does not tend to increase impedances?
a. Increased site preparation
b. Broken/faulty wires
c. Dirty electrodes
d. Poor application - Correct Answer - increased site preparation
Which of these LFF settings is preferred for and EEG channel?
0.5Hz
3.0Hz
10Hz
35Hz - Correct Answer - 0.5Hz
Obstructive apnea is pediatric patients is most often associated with:
a. Idiopathic hypersomnolence
b. Hypothyroidism
c. Hyperactivity
d. Overlap syndrome - Correct Answer - hyperactivity
Which of these factors does not tend to increase impedances?
a. Increased site preparation
b. Broken/faulty wires
c. Dirty electrodes
d. Poor application - Correct Answer - increased site preparation
pg. 2
,3|P a g e
Which of these does not tend to increase leakage current?
a. Shorter power cords
b. Power strips
c. Battery backups
d. Parallel placement of wires - Correct Answer - shorter power cords
The technologist performs physiological calibrations in order to:
a. assess the function of the computer keyboard.
b. assess the function of the electrical bed.
c. verify electrode response to commands.
d. verify the polarity of the electrocardiogram (ECG). - Correct Answer -
c. verify electrode response to commands.
A pattern of cardiac variability often seen in patients with OSA is:
A. Persistent Tachycardia
B. Persistent bradycardia
C. Tachycardia alternating with bradycardia
D. Prolonged QT interval - Correct Answer - Tachycardia alternating
with bradycardia
Accessory equipment such as a CPAP machine or radio placed near the
headbox may cause:
A. A 60-Hz artifact
B. Electrode pops
pg. 3
, 4|P a g e
C. Spurious electrode impedance measures
D. Gap junctions with sparking - Correct Answer - A 60-Hz artifact
The amount of space needed to archive digitized data is most impacted
by:
a. Filter settings
b. Sampling rate
c. Epoch size
d. Sensitivity - Correct Answer - Sampling rate
The measured distance from the nasion to the inion is 36cm. What is the
distance from Cz to Pz?
a. 1.8cm
b. 3.6cm
c. 7.2cm
d. 18cm - Correct Answer - 7.2cm
The most desirable sampling rate for the respiratory airflow signal is:
a. 1 HZ
b. 25Hz
c. 100 Hz
d. 500 Hz - Correct Answer – 100Hz
Real time access to PAP compliance data is an example of:
pg. 4