100% de satisfacción garantizada Inmediatamente disponible después del pago Tanto en línea como en PDF No estas atado a nada 4,6 TrustPilot
logo-home
Examen

FLS Written Exam Questions and Answers 2025

Puntuación
-
Vendido
-
Páginas
13
Grado
A
Subido en
05-06-2025
Escrito en
2024/2025

Laparoscope diameters -Correct Answer 2-10 mm Laparoscope lengths -Correct Answer 30-45 cm Hopkins Rod lens system -Correct Answer light from source to operative field capture and transmit reflected light through the scope and camera couple light sensitive chip in the camera head Decreasing ability for image capture with what factors? -Correct Answer increasing length, increasing angle, and decreasing diameter (ex: 5 mm, 30 degree scope admits less light than a 10 mm, 0 degree scope) Laparoscopic damage susceptibility increases with which factors? -Correct Answer decreasing diameter, increasing length 0 degree laparocope most useful? -Correct Answer when structures most in line with the trocars and in small places (deep pelvis, high mediastinum) 30- and 45-degree laparocope -Correct Answer offer more flexibility/versatility than 0 degree (but need more skills) Troubleshooting smudge on laparoscope? -Correct Answer o Wipe on clean tissue (liver or bowel) o Remove scope and clean with towel o Clean port (newer systems may not need this as no mechanical seal) Warmed/humidified CO2 -Correct Answer decreases postop pain and temperature (compared to gas directly from cylinder) but is not clinically significant in cases 90min High powered light source wattage? -Correct Answer 300 watt Xenon lamp "Picture in Picture" feature -Correct Answer · Helpful if need for intra-op of flexible endoscopy or ultrasound of GI or GU tract, liver ultrasound, hysteroscopy, bile duct exploration o Basically smaller picture in corner to show scope imaging + additional images · Work space decrease in size: Actual pressure higher than set pressure = -Correct Answer o Patient may not be sufficiently relaxed or there is obstruction o Twitch monitors may not represent this well as diaphragm recovery from paralytics differs from skeletal muscles o Mechanical obstruction - closed valve on port, kink, standing on tubing Low pressure and high flow rate = -Correct Answer = leak in system o tubing disconnected from insufflator or port o Valve open—check to make sure valves are cloed o Check all port sites for leaking CO2 o Hollow organs check distension of bowel or bladder catheter Waveforms -Correct Answer coag, cut, blend · Tissue effects: -Correct Answer o Vaporization: 'cut', non-contact, deep injury with minimal lateral damage o Fulguration: 'coag', non-contact, superficial wide injury o Desiccation: either, contact with instrument, deep/penetrating injury --However, 'cut' gives you most thorough desiccation! Harmonic scalpel -Correct Answer · uses mechanical energy, so no dispersive electrode needed o Little lateral injury, good for small vessels · Ligasure -Correct Answer o Bipolar, continuous waveform, measures tissue impedance and shuts down as needed o single application is sufficient! If jewelry cannot be removed, then? -Correct Answer · use gauze and tape to increase contact area and to minimize current concentration · jewelry should not between dispersive electrode and surgical site Current Density = -Correct Answer Current (amps)/Area (cm2) Current Density proportional to -Correct Answer power (the current) Current Density inversely proportional to -Correct Answer o tissue resistance and area squared o Smaller the contact area, greater the current density faster heating o Tip of electrode is small, so contacting a small area of tissue generates high current density dispersive pad (grounding pad) location -Correct Answer o minimal heating and maintains low density -- don't place on hair, bony prominence, or scars which may cause decrease in contact surface area and increase risk of burns

Mostrar más Leer menos
Institución
FLS Written
Grado
FLS Written

Vista previa del contenido

FLS Written



FLS Written Exam Questions and
Answers 2025
Laparoscope diameters -Correct Answer ✔2-10 mm

Laparoscope lengths -Correct Answer ✔30-45 cm

Hopkins Rod lens system -Correct Answer ✔light from source to operative field >>
capture and transmit reflected light through the scope and camera couple >> light
sensitive chip in the camera head

Decreasing ability for image capture with what factors? -Correct Answer ✔increasing
length, increasing angle, and decreasing diameter (ex: 5 mm, 30 degree scope admits
less light than a 10 mm, 0 degree scope)

Laparoscopic damage susceptibility increases with which factors? -Correct Answer
✔decreasing diameter, increasing length

0 degree laparocope most useful? -Correct Answer ✔when structures most in line with
the trocars and in small places (deep pelvis, high mediastinum)

30- and 45-degree laparocope -Correct Answer ✔offer more flexibility/versatility than 0
degree (but need more skills)

Troubleshooting smudge on laparoscope? -Correct Answer ✔o Wipe on clean tissue
(liver or bowel)
o Remove scope and clean with towel
o Clean port (newer systems may not need this as no mechanical seal)

Warmed/humidified CO2 -Correct Answer ✔decreases postop pain and temperature
(compared to gas directly from cylinder)
but is not clinically significant in cases < 90min

High powered light source wattage? -Correct Answer ✔300 watt Xenon lamp

"Picture in Picture" feature -Correct Answer ✔· Helpful if need for intra-op of flexible
endoscopy or ultrasound of GI or GU tract, liver ultrasound, hysteroscopy, bile duct
exploration
o Basically smaller picture in corner to show scope imaging + additional images

· Work space decrease in size: Actual pressure higher than set pressure = -Correct
Answer ✔o Patient may not be sufficiently relaxed or there is obstruction


FLS Written

, FLS Written


o Twitch monitors may not represent this well as diaphragm recovery from paralytics
differs from skeletal muscles
o Mechanical obstruction - closed valve on port, kink, standing on tubing

Low pressure and high flow rate = -Correct Answer ✔= leak in system
o tubing disconnected from insufflator or port
o Valve open—check to make sure valves are cloed
o Check all port sites for leaking CO2
o Hollow organs check distension of bowel or bladder catheter

Waveforms -Correct Answer ✔coag, cut, blend

· Tissue effects: -Correct Answer ✔o Vaporization: 'cut', non-contact, deep injury with
minimal lateral damage
o Fulguration: 'coag', non-contact, superficial wide injury
o Desiccation: either, contact with instrument, deep/penetrating injury
--However, 'cut' gives you most thorough desiccation!

Harmonic scalpel -Correct Answer ✔· uses mechanical energy, so no dispersive
electrode needed
o Little lateral injury, good for small vessels

· Ligasure -Correct Answer ✔o Bipolar, continuous waveform, measures tissue
impedance and shuts down as needed
o single application is sufficient!

If jewelry cannot be removed, then? -Correct Answer ✔· use gauze and tape to increase
contact area and to minimize current concentration
· jewelry should not between dispersive electrode and surgical site

Current Density = -Correct Answer ✔Current (amps)/Area (cm2)

Current Density proportional to -Correct Answer ✔power (the current)

Current Density inversely proportional to -Correct Answer ✔o tissue resistance and area
squared
o Smaller the contact area, greater the current density >> faster heating
o Tip of electrode is small, so contacting a small area of tissue generates high current
density

dispersive pad (grounding pad) location -Correct Answer ✔o minimal heating and
maintains low density
-- don't place on hair, bony prominence, or scars which may cause decrease in contact
surface area and increase risk of burns



FLS Written

Escuela, estudio y materia

Institución
FLS Written
Grado
FLS Written

Información del documento

Subido en
5 de junio de 2025
Número de páginas
13
Escrito en
2024/2025
Tipo
Examen
Contiene
Preguntas y respuestas

Temas

$15.49
Accede al documento completo:

100% de satisfacción garantizada
Inmediatamente disponible después del pago
Tanto en línea como en PDF
No estas atado a nada

Conoce al vendedor

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
PossibleA Chamberlain College Of Nursing
Seguir Necesitas iniciar sesión para seguir a otros usuarios o asignaturas
Vendido
999
Miembro desde
4 año
Número de seguidores
650
Documentos
12879
Última venta
2 días hace
POSSIBLEA QUALITY UPDATED EXAMS

Choose quality study materials for nursing schools to ensure success in your studies and future career. "Welcome to PossibleA - your perfect study assistant! Here you will find Quality sheets, study materials, exams, quizzes, tests, and notes to prepare for exams and study successfully. Our store offers a wide selection of materials on various subjects and difficulty levels, created by experienced teachers and checked for quality. Our quality sheets are an easy and quick way to remember key points and definitions. And our study materials, tests, and quizzes will help you absorb the material and prepare for exams. Our store also has notes and lecture summaries that will help you save time and make the learning process more efficient.

Lee mas Leer menos
3.9

143 reseñas

5
74
4
25
3
21
2
1
1
22

Recientemente visto por ti

Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes