100% de satisfacción garantizada Inmediatamente disponible después del pago Tanto en línea como en PDF No estas atado a nada 4.2 TrustPilot
logo-home
Examen

Fluoroscopy Test - Image Production Exam Questions with Complete Answers

Puntuación
-
Vendido
-
Páginas
7
Grado
A+
Subido en
29-05-2025
Escrito en
2024/2025

Fluoroscopy Test - Image Production Exam Questions with Complete Answers

Institución
Fluoroscopy
Grado
Fluoroscopy









Ups! No podemos cargar tu documento ahora. Inténtalo de nuevo o contacta con soporte.

Escuela, estudio y materia

Institución
Fluoroscopy
Grado
Fluoroscopy

Información del documento

Subido en
29 de mayo de 2025
Número de páginas
7
Escrito en
2024/2025
Tipo
Examen
Contiene
Preguntas y respuestas

Temas

Vista previa del contenido

Fluoroscopy Test - Image Production
Exam Questions with Complete
Answers

Atomic brightness control (ABC) maintains the brightness of the image by varying
______ - ANSWER-kVp and mA

The time it takes to turn on the digital fluoroscopy x-ray tube and reach the selected
mA and kVp is the ____ time - ANSWER-interrogation

Digital fluoroscopy with a charge-couples device has lower _____ and higher _____
than conventional fluoroscopy - ANSWER-patient dose, light sensitivity

A principal advantage of digital fluoroscopy is the _____ - ANSWER-image
subtraction

Digital fluoroscopic dose rate is lower than that for continuous analog fluoroscopy -
ANSWER-True

The source-to-skin distance limit for stationary fluoroscopic unit is _____ cm -
ANSWER-38

When performing a moblie fluoroscopic procedure, to reduce the radiation exposure
to the patient, the radiographer must use a minimal source-skin distance of _____
cm (_______ inches) - ANSWER-30; 12

How is brightness gain determined - ANSWER-product of minification and flux gain

Image contrast can be improved with ____ techniques - ANSWER-Subtraction

The image captured by FPIR is ________ compared to the ______ image of
traditional fluoroscopy - ANSWER-square, circular

The image intensifier's input phosphor differs from the output phosphor in what way -
ANSWER-is much larger

The time it takes to turn on the digital fluoroscopy x-ray tube and reach the selected
mA and kVp is the _____ time - ANSWER-interrogation

The total dose rate at the table top during fluoroscopy cannot exceed - ANSWER-10
R per minute

What device is required with a direct digital type of fluoroscopy system - ANSWER-
flat detector

, What happens when a multifocus image intensifier tube is operated in the
magnification mode - ANSWER-the electron focal point is closer to the input
phosphor

When using a 25/17/12 image intensifier tube, which of the following is true: -
ANSWER-the field of view is largest in the 25 cm mode

Which of the following describes the fluoroscopic system designed to maintain a
constant image brightness - ANSWER-automatic brightness control

Which of the following modes of a trifield image intensifier will result in the highest
patient dose - ANSWER-its 12 inch mode

If using a dynamic flat panel detector in fluoroscopy, the image intensifier is no
longer needed - ANSWER-True

The FPIR signal when used in DF is already digital - ANSWER-True

What is the minification gain for an image intensification tube with an input screen
diameter of 5 inches and an output diameter of 1 inch - ANSWER-25

What is the total brightness gain for an image intensification tube with a minification
gain of 25 and a fluc gain of 50 - ANSWER-1250

An image intensifier tube is identified by the diameter of its - ANSWER-input
phosphor

Brightness gain is typically in the range of _____. - ANSWER-5000-30000

During fluoroscopic image intensification, the primary x-ray beam exits the patient
and strikes the ____ of the image intensifier. - ANSWER-input screen

electrons hit the ___ after exiting the anode - ANSWER-output phosphor

Electrostatic lenses are actually _____ - ANSWER-charged electrodes

Electrostatic lenses are used to accelerate and focus - ANSWER-electrons

The image intensifier improved fluoroscopy by increasing image - ANSWER-
brightness

The input screen absorbs _____ and emits _____ - ANSWER-xray photons, light
photons

The kinetic energy of photoelectrons in the image intensifier is greatly increased by
the - ANSWER-potential difference across the tube

The number of light photons emitted within the image intensifier is ______ to the
amount of xray photons exiting the patient - ANSWER-directly proportional
$18.49
Accede al documento completo:

100% de satisfacción garantizada
Inmediatamente disponible después del pago
Tanto en línea como en PDF
No estas atado a nada


Documento también disponible en un lote

Conoce al vendedor

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
lectknancy Boston University
Seguir Necesitas iniciar sesión para seguir a otros usuarios o asignaturas
Vendido
284
Miembro desde
2 año
Número de seguidores
28
Documentos
26008
Última venta
1 día hace

3.6

57 reseñas

5
23
4
10
3
11
2
3
1
10

Recientemente visto por ti

Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes