ASQ Calibration Technician Certification
Practice Exam-Graded A
The SI unit of length is the - correct Answer-Meter
The Pascal, the SI unit of pressure, is - correct Answer-A derived SI unit
Which SI multipliers have abbreviations in capital letters? - correct Answer-Those with
factors > 1000
The speed of light is considered to be a constant - correct Answer-When used for
standards work
The melting point of Gallium is a fixed point in the ITS-90 temperature system. The
uncertainty of the value of this temperature is - correct Answer-Zero
A certificate for a traceable calibration must show evidence to support - correct Answer-
The existence of a traceability path is needed for an audit
The realization of a temperature fixed point is - correct Answer-A physical object such
as a freezing point cell
Traceability is required when operating under ISO 9001 and related standards - correct
Answer-At every step of the calibration pyramid below national measurement standards
The ACME company's calibration procedure for their model VU-1 sound level meter
calls for the exclusive use of ACME sound level calibrator model SLC-1. A technician
has a rush job to calibrate a VU-1, but the SLC-1 is back at the factory in Europe for
repair and calibration. He/she should - correct Answer-Stop work and notify the
customer that the calibration cannot be completed until the proper IM&TE is available
In order to calibrate a digital ammeter, it is necessary to have - correct Answer-A source
or generator of current
The base SI unit for electrical quantities is - correct Answer-Ampere
One gram of oxygen gas is released into a box 10 cm on a side. What is the SI density
of the gas inside the box when equilibrium is reached? - correct Answer-1.0 kg/m^3
One million nanofarads is the same as - correct Answer-One millifarad
,Scales and balances should be calibrated after they are moved because - correct
Answer-Gravity is different in different places and internal mechanisms may have been
affected
A triple point of water cell is - correct Answer-A device for sourcing temperature
Substitution of standards - correct Answer-May be done if technically feasible and not
forbidden by procedure
Which of the following measuring systems is a resistance-temperature device (RTD)? -
correct Answer-Thermistor
Having traceable reference standards and providing traceable calibrations for a
laboratory's customers is necessary when - correct Answer-These customers need to
provide traceability to their customers
An optical pyrometer measures temperature by determining - correct Answer-The color
of the light from a hot body
An intrinsic standard - correct Answer-Will realize the unit for which it is a reference
when operated properly
The largest single contributor to the uncertainty of dimensional measurements is -
correct Answer-Temperature
The SI unit for time is based on - correct Answer-An atomic vibration
Which of the following notations accepted for use with the SI is preferred for the volume
of a gallon of milk? - correct Answer-3.785 L (1 U.S. gallon)
Carl Sagan, on his television series, might have said "Thank you for spending these
billions and billions of nanoseconds with me." How long a time interval is 1800 billion
ns? Note - use the U.S. definition of billion, e.g. 10^9 - correct Answer-Half an hour
Gravitational corrections are always required for - correct Answer-Measurements in
which a weight opposes a spring force
Which of the following temperature measuring devices requires either an actual or a
simulated reference temperature? - correct Answer-Thermocouple
Crash dummies used to simulate the effects of adults and children during testing of car
and truck collisions are an example of a - correct Answer-Consensus standard
Electrical sensors used to measure temperature almost always require - correct
Answer-Linearization of the temperature - electrical relationship
, A substitute standard has replaced the usual reference standard and the term "limited
calibration" is used on the UUT. Which of the following explanations would NOT be
expected as an explanation of the limited calibration? - correct Answer-The substitute
was used for a limited time
The most common working standard for dimensional analysis is - correct Answer-The
gage block
Measurement of humidity can be accomplished - correct Answer-With a mirror whose
temperature can be controlled
A laboratory that uses an intrinsic standard may not simply use it without verification,
but must also - correct Answer-Participate in interlaboratory comparisons or proficiency
testing for that parameter
The relationship between the freezing point of water and the triple point of water at
standard atmospheric pressure is - correct Answer-The triple point is warmer
The SI unit used by torque wrenches is - correct Answer-N*m
What ratio of attenuator would be required to reduce a 1.0 mW signal to a 1.0 fW
signal? - correct Answer-1 000 000 000 000 :1
The best reason that the local acceleration of gravity, little g, needs to be known by a
laboratory is because - correct Answer-It can affect many calibration results
When using saturated salt solutions to generate (source) humidity, it is important to
know the - correct Answer-Temperate
A client has submitted a 0 - 250 ft-lb, +- 5% of full scale, torque wrench to a lab for
calibration in the clockwise direction. The lab's 0 - 600 ft-lb, 0.25% of reading, clockwise
and counter-clockwise, electronic calibrator is out of order. The lab also owns a spring-
and-needle torque clockwise calibrator that is working and is in calibration. However, it
is 0 - 1000 ft-lb, 1% of reading 2 ft-lb. Can this calibrator be used instead? - correct
Answer-Yes, it is adequate for this job
Which of the following parameters can be converted to relative humidity using a table or
calculation? - correct Answer-Dew point
ANSI/NCSL Z540.3 requires that calibration procedures shall contain generic
descriptions of the IM&TE and reference standards to be used so that - correct Answer-
The technician has all the required information to make a substitution decision
A certificate of calibration issued by an NMI - correct Answer-May be considered
traceable without further evidence
Practice Exam-Graded A
The SI unit of length is the - correct Answer-Meter
The Pascal, the SI unit of pressure, is - correct Answer-A derived SI unit
Which SI multipliers have abbreviations in capital letters? - correct Answer-Those with
factors > 1000
The speed of light is considered to be a constant - correct Answer-When used for
standards work
The melting point of Gallium is a fixed point in the ITS-90 temperature system. The
uncertainty of the value of this temperature is - correct Answer-Zero
A certificate for a traceable calibration must show evidence to support - correct Answer-
The existence of a traceability path is needed for an audit
The realization of a temperature fixed point is - correct Answer-A physical object such
as a freezing point cell
Traceability is required when operating under ISO 9001 and related standards - correct
Answer-At every step of the calibration pyramid below national measurement standards
The ACME company's calibration procedure for their model VU-1 sound level meter
calls for the exclusive use of ACME sound level calibrator model SLC-1. A technician
has a rush job to calibrate a VU-1, but the SLC-1 is back at the factory in Europe for
repair and calibration. He/she should - correct Answer-Stop work and notify the
customer that the calibration cannot be completed until the proper IM&TE is available
In order to calibrate a digital ammeter, it is necessary to have - correct Answer-A source
or generator of current
The base SI unit for electrical quantities is - correct Answer-Ampere
One gram of oxygen gas is released into a box 10 cm on a side. What is the SI density
of the gas inside the box when equilibrium is reached? - correct Answer-1.0 kg/m^3
One million nanofarads is the same as - correct Answer-One millifarad
,Scales and balances should be calibrated after they are moved because - correct
Answer-Gravity is different in different places and internal mechanisms may have been
affected
A triple point of water cell is - correct Answer-A device for sourcing temperature
Substitution of standards - correct Answer-May be done if technically feasible and not
forbidden by procedure
Which of the following measuring systems is a resistance-temperature device (RTD)? -
correct Answer-Thermistor
Having traceable reference standards and providing traceable calibrations for a
laboratory's customers is necessary when - correct Answer-These customers need to
provide traceability to their customers
An optical pyrometer measures temperature by determining - correct Answer-The color
of the light from a hot body
An intrinsic standard - correct Answer-Will realize the unit for which it is a reference
when operated properly
The largest single contributor to the uncertainty of dimensional measurements is -
correct Answer-Temperature
The SI unit for time is based on - correct Answer-An atomic vibration
Which of the following notations accepted for use with the SI is preferred for the volume
of a gallon of milk? - correct Answer-3.785 L (1 U.S. gallon)
Carl Sagan, on his television series, might have said "Thank you for spending these
billions and billions of nanoseconds with me." How long a time interval is 1800 billion
ns? Note - use the U.S. definition of billion, e.g. 10^9 - correct Answer-Half an hour
Gravitational corrections are always required for - correct Answer-Measurements in
which a weight opposes a spring force
Which of the following temperature measuring devices requires either an actual or a
simulated reference temperature? - correct Answer-Thermocouple
Crash dummies used to simulate the effects of adults and children during testing of car
and truck collisions are an example of a - correct Answer-Consensus standard
Electrical sensors used to measure temperature almost always require - correct
Answer-Linearization of the temperature - electrical relationship
, A substitute standard has replaced the usual reference standard and the term "limited
calibration" is used on the UUT. Which of the following explanations would NOT be
expected as an explanation of the limited calibration? - correct Answer-The substitute
was used for a limited time
The most common working standard for dimensional analysis is - correct Answer-The
gage block
Measurement of humidity can be accomplished - correct Answer-With a mirror whose
temperature can be controlled
A laboratory that uses an intrinsic standard may not simply use it without verification,
but must also - correct Answer-Participate in interlaboratory comparisons or proficiency
testing for that parameter
The relationship between the freezing point of water and the triple point of water at
standard atmospheric pressure is - correct Answer-The triple point is warmer
The SI unit used by torque wrenches is - correct Answer-N*m
What ratio of attenuator would be required to reduce a 1.0 mW signal to a 1.0 fW
signal? - correct Answer-1 000 000 000 000 :1
The best reason that the local acceleration of gravity, little g, needs to be known by a
laboratory is because - correct Answer-It can affect many calibration results
When using saturated salt solutions to generate (source) humidity, it is important to
know the - correct Answer-Temperate
A client has submitted a 0 - 250 ft-lb, +- 5% of full scale, torque wrench to a lab for
calibration in the clockwise direction. The lab's 0 - 600 ft-lb, 0.25% of reading, clockwise
and counter-clockwise, electronic calibrator is out of order. The lab also owns a spring-
and-needle torque clockwise calibrator that is working and is in calibration. However, it
is 0 - 1000 ft-lb, 1% of reading 2 ft-lb. Can this calibrator be used instead? - correct
Answer-Yes, it is adequate for this job
Which of the following parameters can be converted to relative humidity using a table or
calculation? - correct Answer-Dew point
ANSI/NCSL Z540.3 requires that calibration procedures shall contain generic
descriptions of the IM&TE and reference standards to be used so that - correct Answer-
The technician has all the required information to make a substitution decision
A certificate of calibration issued by an NMI - correct Answer-May be considered
traceable without further evidence