SPI PRACTICE QUESTIONS 2024/2025
_________ is a redirection of US in many directions as a result of a *rough* boundary between two media - RIGHT ANSWER -scattering
___________ resolution is determined by SPL - RIGHT ANSWER -radial or axial
______________ is the distance covered by one cycle - RIGHT ANSWER -wavelength
______________ is the time to complete one cycle - RIGHT ANSWER -period
A circular hyperechoic structure is located at a depth of 5 cm. The structure has abnormally low attenuation and the speed in the mass is 1,400 m/s. Which of the following is most likely to appear beneath the structure? - RIGHT ANSWER -reflectors beneath the structure appear abnormally deep in the image
A Doppler exam is performed with a 5 MHz probe and a PRF of 15 kHz. Which of these Doppler shifts will create aliasing? - RIGHT ANSWER -14 kHz
A hydrophone is used to measure: - RIGHT ANSWER -intensity (or frequency)
A pulse is emitted by a transducer and is traveling in soft tissue. The go-return-time, or time-of-flight, of a sound pulse is 130 microseconds. What is the reflector depth? - RIGHT ANSWER -10 cm
A quality assurance program: - RIGHT ANSWER -is needed by all labs
A shadow appears on an ultrasound scan. Which of the following statements is true about the shadow artifact? - RIGHT ANSWER -it appears as hypoechoic
A shadow appears on an ultrasound scan. Which of the following statements is true about the structure immediately above the shadowing artifact? - RIGHT ANSWER -attenuation properties are too high
A single element transducer is focused: - RIGHT ANSWER -by the manufacturer
A sound beam is created by a single, disc-shaped crystal. Which of these beams has the shallowest focus? - RIGHT ANSWER -2 MHz, 9 mm diameter
A sound wave is traveling in the body and propagating from one form of soft tissue to another. For example, sound is moving from muscle to blood. What percentage of the sound wave is most likely to be reflected at the muscle-blood boundary? - RIGHT ANSWER -.5%
A sound wave with a frequency of 5 MHz is travelling in soft tissue. What is the wavelength? - RIGHT ANSWER -.3 mm
A sound wave's frequency is 10 MHz. The wave is traveling in soft tissue. What is its wavelength? - RIGHT ANSWER -.15 mm
A system which is not working well now does not display clear images in the far field. This may be a problem with: - RIGHT ANSWER -transducer, receiver, monitor, or scan converter
A transducer converts __________ from one form to another - RIGHT ANSWER -energy
A transducer with more than one active element is called: - RIGHT ANSWER -array
According to AIUM, there have been no confirmed bioeffects below intensities of ______ watts per square centimeter SPTA - RIGHT ANSWER -.01
Aliasing is an artifact observed when performing Doppler ultrasonography. Which of the following signals is most likely to alias? - RIGHT ANSWER -7 MHz with a PRF of 3 kHz
All of the following are limitations of analog scan converts EXCEPT: - RIGHT ANSWER -low spatial resolution
All of the following are primarily used to change the overall brightness of an entire image EXCEPT: - RIGHT ANSWER -compensation
All of the following are true EXCEPT: - RIGHT ANSWER -two waves with identical frequencies must interfere constructively
All of the following are true EXCEPT: - RIGHT ANSWER -two waves with identical frequencies must interfere constructively
All of the following are true of normal cavitation EXCEPT: - RIGHT ANSWER -harmful bioeffects are commonly observed
All of the following are true regarding temporal resolution EXCEPT: - RIGHT ANSWER -decreased temporal resolution is associated with shorter pulse duration
All of the following artifacts place the erroneous reflection below the true reflector EXCEPT: - RIGHT ANSWER -side lobe
All of the following describe dynamic range EXCEPT: logarithmic, comparison, relative, absolute, ratio - RIGHT ANSWER -absolute
All of the following regarding the hemodynamics of a stenosis are true EXCEPT: - RIGHT ANSWER -pressure is lowest downstream
All of the following result from the use of backing material in a transducer EXCEPT: - RIGHT ANSWER -high sensitivity
All of the following statements regarding in-vitro bioeffects are true EXCEPT: - RIGHT ANSWER -in-vitro bioeffects conclusions are clinically relevant
All of the following will improve temporal resolution EXCEPT: - RIGHT ANSWER -increased line density
An image is too dark. Which technique is best to correct this problem? - RIGHT ANSWER -increase receiver gain
An individual is standing. The blood pressure generated by the heart is 120 mmHg. What is the hydrostatic pressure at the ankle? - RIGHT ANSWER -100 mmHg
An ultrasound system is imaging to a depth of 10 cm. Which of the following will be different on a single image at depths of 2 cm, 5 cm, and 10 cm? - RIGHT ANSWER -lateral resolution
An ultrasound system is operating in the harmonic mode. The image on the display of the ultrasound system was created using sound reflections with a frequency of 6 MHz. What is the frequency of sound emitted by the system's transducer? - RIGHT ANSWER -3 MHz
An ultrasound system is set at 0 dB and is transmitting at full intensity. What is the output power when the system is transmitting at 10% of full intensity? - RIGHT ANSWER --10 dB
An ultrasound system is set at 0 dB and is transmitting at full intensity. What is the output power when the system is transmitting at 50% of full intensity? - RIGHT ANSWER --3 dB
An ultrasound system's display is too bright. Which of the following techniques is best to correct this problem? - RIGHT ANSWER -decrease output power
An US machine has a digital scan converter, but has an analog input signal. The machine needs to have a: - RIGHT ANSWER -analog to digital converter
An US wave travels through two media and has attenuation of 3 dB in the first and 5 dB in the second. What is the total attenuation? - RIGHT ANSWER -8 dB
As a result of ______________, the propagation speed increases - RIGHT ANSWER -decreasing compressibility and density
As the impedance of two media become vastly different , the: - RIGHT ANSWER -reflection increases
Assuming a constant frequency, what happens if the diameter of an unfocused circular transducer is increased? - RIGHT ANSWER -the beam width in the near field is increased
At the focus, the beam diameter is __________ the transducer diameter - RIGHT ANSWER -half
Attenuation of US in soft tissue is closest to: - RIGHT ANSWER -.6 dB/cm/MHz
Biologic effects have not been documented below what intensity? (SPTA) - RIGHT ANSWER -100 mW/cm2
Changing which of the following would not cause any change in a hard copy image output? - RIGHT ANSWER -display brightness and contrast
Circumferences may be measured in the unit of: - RIGHT ANSWER -cm
Compression is used to: - RIGHT ANSWER -decrease the range of amplitudes
Decimal uses a base of 10, binary uses a base of: - RIGHT ANSWER -2
Depth calibration of a machine measures 100 mm spaced wires to be 90 mm apart. Scanning reflectors 50 mm apart, what will the machine calculate the distance to be? - RIGHT ANSWER -45 mm
Enhancement artifact appears on an ultrasound scan. Which of the following statements is true about the structure immediately above the artifact? - RIGHT ANSWER -attenuation properties are too low
Enhancement, multipath and side lobes result in: - RIGHT ANSWER -artifact
For Doppler, which incident angle results in NO shift? - RIGHT ANSWER -90 degrees
For soft tissue, the approximate attenuation coefficient in dB/cm is equal to one-half _________ in _________ - RIGHT ANSWER -frequency, MHz
From a safety standpoint, which one of the following methods is best? - RIGHT ANSWER -low transmit output and high receiver gain
How does frequency compounding improve image quality? - RIGHT ANSWER -reduces speckle
How is range ambiguity artifact eliminated? - RIGHT ANSWER -lower the PRF
How many bits are needed to represent 1024 shades of gray? - RIGHT ANSWER -10
How many bits are needed to represent 16 shades of grey - RIGHT ANSWER -4
How many cm are 60 mm? - RIGHT ANSWER -6
How many data bits are needed to represent 64 gray shades? - RIGHT ANSWER -6
How much bigger is a billion than a thousand? - RIGHT ANSWER -1,000,000
How much time is required for a sound pulse to travel round trip to and from a reflector that is 2 cm deep in soft tissue? - RIGHT ANSWER -26 microseconds
Identify two advantages of coded excitation: - RIGHT ANSWER -improved contrast resolution and penetration
If 300,000 cycles occur in a second, what is the wave's frequency? - RIGHT ANSWER -.3 MHz
If a 5 MHz sound wave is traveling in the tissues listed below, which will have the longest wavelength? - RIGHT ANSWER -bone
If a 5 MHz sound wave is traveling in the tissues listed below, which will have the shortest wavelength? - RIGHT ANSWER -air
If a manufacturer changed the lines per frame and kept the imaging depth the same, what else would change? - RIGHT ANSWER -frame rate
If a reflector is moved twice as far away from the transducer, how will this affect the time-of-flight of a sound pulse? - RIGHT ANSWER -double it
If a wave's amplitude is doubled, what happens to the power? - RIGHT ANSWER -quadrupled
If the frame rate increases and the lines per frame is unchanged, what else must happen? - RIGHT ANSWER -depth decreases
If the frequency is decreased, the numerical value of the radial resolution is: - RIGHT ANSWER -increased
If the frequency of an US wave is doubled, what happens to the period? - RIGHT ANSWER -halved
If the frequency of US is increased from 0.77 MHz to 1.54 MHz, what happens to the wavelength? - RIGHT ANSWER -halved
If the frequency of US is increased from 0.77MHz to 1.54MHz, what happens to the propagation speed? - RIGHT ANSWER -remains the same
If the intensity is increased by 3 dB, it: - RIGHT ANSWER -doubles
If the intensity of a sound beam remains unchanged while the beam area is reduced in half, what has happened to the power? - RIGHT ANSWER -halved
If the lines per frame is increased while the imaging depth is unchanged then: - RIGHT ANSWER -the frame rate decreases
If the SPL is 10 mm, what is the axial resolution? - RIGHT ANSWER -.5 cm
If the transducer diameter increases, the lateral resolution at its smallest dimension is: - RIGHT ANSWER -increased
If we increase frequency, the near zone length is: - RIGHT ANSWER -increased
If we increase the transducer diameter, the beam diameter in the far zone is: - RIGHT ANSWER -decreased
If wires in an AIUM test object do not appear in the same place as on the display, what is amiss? - RIGHT ANSWER -registration accuracy
Images on videotapes are stored using: - RIGHT ANSWER -magnetism
Impedance is the _________ of _________ and _________ - RIGHT ANSWER -product, propagation speed, density
Impedance is the product of __________ and __________ - RIGHT ANSWER -density and propagation speed
Improving the ability to measure the maximum velocity with Doppler is accomplished with all of the following EXCEPT: - RIGHT ANSWER -increasing the transducer's frequency
In a B-mode display, which of the following axis are used for the amplitude of the reflection? - RIGHT ANSWER -z-axis
In a real-time scanner, which of these choices determines the maximum frame rate? - RIGHT ANSWER -depth of view
In an air-tissue interface, what percentage of the US is reflected? - RIGHT ANSWER -near 100%
In B-mode, which input is used with the amplitude of reflection? - RIGHT ANSWER -z-axis
In soft tissue:
Distance traveled = __________ x round trip time - RIGHT ANSWER -propagation speed
In which of these biologic tissues will sound waves propagate most quickly? - RIGHT ANSWER -tendon
In which of these media will sound travel the slowest? - RIGHT ANSWER -high density, low stiffness
In which way is an annular phase array transducer different than a linear phased array transducer? - RIGHT ANSWER -the annular is steered mechanically
Lateral resolution can be improved by: - RIGHT ANSWER -focusing
List the following in increasing order: deci, kilo, micro, mega, milli - RIGHT ANSWER -micro, milli, deci, kilo, mega
Multiple reflections that are equally spaced are called: - RIGHT ANSWER -reverberations
Post processing occurs when the image: - RIGHT ANSWER -entered the A/D converter but before display
Power of sound has units of: - RIGHT ANSWER -watts
Pre-processing occurs: - RIGHT ANSWER -when the patient is scanned
Propagation speed = ______________ x wavelength - RIGHT ANSWER -frequency
Put in decreasing order of propagation speed: gas, liquid, solid - RIGHT ANSWER -solid, liquid, gas
Put in decreasing order: deci, deca, micro, centi - RIGHT ANSWER -deca, deci, centi micro
Put in increasing order of propagation speed: bone, water, air - RIGHT ANSWER -air, water, bone
Quality assurance studies should be performed: - RIGHT ANSWER -routinely
Rayl is the unit of: - RIGHT ANSWER -impedance
Rectification and smoothing are components of: - RIGHT ANSWER -demondulation
Refraction only occurs if there are: - RIGHT ANSWER -oblique incidence and different propagation speeds
Reject is used to: - RIGHT ANSWER -adjust the appearance of lowly reflective objects
Snell's law describes the physics of: - RIGHT ANSWER -refraction
Sound intensity is decreased by 75%. How many decibels of attenuation is that? - RIGHT ANSWER -6 dB
Sound is traveling from medium 1 to medium 2. Propagation speeds of M1 is 1,600 m/s and M2 is 1,500 m/s. There is oblique incidence - RIGHT ANSWER -angle of incidence = angle of reflection
Sound waves are: - RIGHT ANSWER -longitudinal, mechanical
T/F: A sonographer's education ends when they become registered - RIGHT ANSWER -false
T/F: A sound wave is made of compressions and rarefactions - RIGHT ANSWER -true
T/F: Axial resolution is affected by focusing - RIGHT ANSWER -false
T/F: Doppler shifts ALWAYS occur is the sound source and receiver are in motion - RIGHT ANSWER -false
T/F: Doppler shifts are always created when the source and receiver are in motion relative to each other - RIGHT ANSWER -false
T/F: Gray scale can be changed by the sonographer - RIGHT ANSWER -true
T/F: Increasing the frequency increases the penetration depth - RIGHT ANSWER -false
T/F: Lateral resolution is also called azimuthal - RIGHT ANSWER -true
T/F: Lateral resolution is also called longitudinal - RIGHT ANSWER -false
T/F: Shadowing artifact is unrelated to the speed of sound in a medium - RIGHT ANSWER -true
T/F: Shadowing may result from high amounts of reflection of US energy - RIGHT ANSWER -true
T/F: Sonographers cannot affect a patient's exposure to acoustic energy - RIGHT ANSWER -false
T/F: The sonographer CANNOT change, directly or indirectly, the duty factor of an US machine - RIGHT ANSWER -false
T/F: There have NOT been biological effects from US found in the labratory - RIGHT ANSWER -false
T/F: With normal incidence, the angle is 180 degrees. - RIGHT ANSWER -false
T/F: With right angle incidence, reflections are always generated at a boundary if the propagation speeds of the media are different - RIGHT ANSWER -false
The angle between two media is 56 degrees. This is called: - RIGHT ANSWER -oblique incidence
The angle between two media is 90 degrees. What is this called? - RIGHT ANSWER -orthogonal incidence
The angle of incidence of an US beam is perpendicular to an interface. The two media have the SAME propagation speeds. What process cannot occur? - RIGHT ANSWER -refraction
The consolidation of many images into a single frame best describes which of the following? - RIGHT ANSWER -persistence
The creation of an image by scanning from different angles best describes: - RIGHT ANSWER -spatial compounding
The Curie temperature has which effect? - RIGHT ANSWER -the ferroelectric material is depolarized
The dB is defined as the __________ of two intensities - RIGHT ANSWER -ratio
The distance to a target is doubled. The time-of-flight for a pulse to travel to the target and back is: - RIGHT ANSWER -doubled
The Doppler shift is a measure of: - RIGHT ANSWER -difference between incident & reflected frequencies
The Doppler shift is the __________ of two frequencies - RIGHT ANSWER -difference
The duty factor for CW ultrasound is: - RIGHT ANSWER -1.0
The dynamic range is the ratio of the smallest to the largest ____________ that a system can process without distortion - RIGHT ANSWER -amplitude
The effects of soft tissue on ultrasound are called: - RIGHT ANSWER -acoustic propagation properties
The fill-in of anechoic structures such as cysts is known as all of the following EXCEPT: - RIGHT ANSWER -ghosting artifact
The frequency closest to the lower limit of ultrasound is: - RIGHT ANSWER -15,000 Hz
The Fresnel zone is the: - RIGHT ANSWER -near zone
The fundamental frequency of a transducer is 2.5 MHz. What is the second harmonic frequency? - RIGHT ANSWER -5 MHz
The general mechanism which creates harmonics is called: - RIGHT ANSWER -non-linear behavior
The more bits per pixel: - RIGHT ANSWER -the more shades of gray
The more pixels per inch: - RIGHT ANSWER -better spatial resolution
The most likely amount of reflection at a boundary between soft tissues is: - RIGHT ANSWER -1%
The most typical amount of transmission at a boundary in biologic media is: - RIGHT ANSWER -99%
The process of reducing the difference between the smallest and largest voltages is called: - RIGHT ANSWER -compression
The production of a voltage from deformation is called: - RIGHT ANSWER -piezoelectricity
The propagation speed of US in the AIUM test object is: - RIGHT ANSWER -1.54 mm/microseconds
The range equation relates distance from the reflector to: - RIGHT ANSWER -time-of-flight and propagation speed
The region from the transducer to the minimum beam cross-sectional area is the: - RIGHT ANSWER -fresnel zone
The relative output of an US instrument is calibrated in dB and the operator increases the output by 60 dB. The beam intensity is increased by which of the following? - RIGHT ANSWER -1,000,000
The speed of sound in soft tissue is closest to: - RIGHT ANSWER -1,500 m/s
The time from the beginning of a pulse until its end is ______________? - RIGHT ANSWER -pulsed duration
The units of attenuation are ______; this is a ______________ scale - RIGHT ANSWER -dB, logarithmic
The units of PRF are: - RIGHT ANSWER -per minute
To measure the US beam intensity, use a: - RIGHT ANSWER -calibrated hydrophone
Traveling through the same medium, if the frequency is doubled, the propagation speed is: - RIGHT ANSWER -unchanged
Traveling through the same medium, if the frequency is doubled, the wavelength is: - RIGHT ANSWER -halved
Two objects start at the same point. One object moves towards the west at 5 miles per hour. The other object moves towards the west at 15 miles per hour. A sound wave is sent from the first object to the second to measure a Doppler frequency. Which of the following statements is true? - RIGHT ANSWER -the doppler frequency will be negative
Under which circumstance is cavitation most likely to occur? - RIGHT ANSWER -low frequency and high signal amplitude
Under which condition is axial resolution improved? - RIGHT ANSWER -decreased pulse duration
Under which of the following conditions is axial resolution improved? - RIGHT ANSWER -decreased SPL
Under which of the following conditions is lateral resolution improved? - RIGHT ANSWER -decreased beam diameter
Volume has which of the following units? - RIGHT ANSWER -cm^3
What are attenuation's three components? - RIGHT ANSWER -scattering, reflection, absorption
What are the units of intensity? - RIGHT ANSWER -watts/cm2
What are the units of longitudinal resolution? - RIGHT ANSWER -feet
What are typical clinical Doppler frequencies? - RIGHT ANSWER -1-10,000 Hz
What artifact has a grainy appearance and is caused by the interference effects of scattered sound? - RIGHT ANSWER -speckle
What best describes the features of a sound wave and medium that determine the speed of sound? - RIGHT ANSWER -density and stiffness
What can pulsed Doppler measure that continuous wave cannot? - RIGHT ANSWER -location
What component of a transducer changes electrical to mechanical and mechanical back to electrical energy? - RIGHT ANSWER -PZT
What component of the US unit contains the memory bank? - RIGHT ANSWER -scan converter
What determines the center frequency of a transducer? - RIGHT ANSWER -thickness
What determines the frequency of a sound beam from a pulsed transducer? - RIGHT ANSWER -PZT thickness
What determines the spatial resolution of a monitor? - RIGHT ANSWER -the number of lines per frame
What determines whether reflection will occur when a sound beam strikes a boundary at normal incidence? - RIGHT ANSWER -impedance differences
What does a hydrophone measure? - RIGHT ANSWER -acoustic output of an ultrasound imaging system
What Doppler angle is most inaccurate? - RIGHT ANSWER -60-75 degrees
What happens to the speed of sound in a medium when the bulk modulus of the medium decreases? - RIGHT ANSWER -slower
What happens to the speed of sound in a medium when the bulk modulus of the medium increases? - RIGHT ANSWER -faster
What happens to venous blood flow to the heart during expiration? - RIGHT ANSWER -decreased
What has the greatest influence upon color jet size? - RIGHT ANSWER -color doppler gain settings
What increases patient exposure? - RIGHT ANSWER -exam time
What information does Doppler shift provide? - RIGHT ANSWER -velocity
What is another name for longitudinal resolution? - RIGHT ANSWER -axial or depth
What is displayed on the x-axis of an M-mode image? - RIGHT ANSWER -time
What is displayed on the y-axis of an M-mode image? - RIGHT ANSWER -time-of-flight
What is NOT associated with PACS? - RIGHT ANSWER -analog replay
What is NOT related to CW wave transducers? - RIGHT ANSWER -wide bandwidth
What is not the responsibility of the sonographer? - RIGHT ANSWER -adjust converter matrix
What is NOT true regarding normal cavitation? - RIGHT ANSWER -harmful bioeffects are commonly observed
What is related to frequency to the 4th power? - RIGHT ANSWER -rayleigh scattering
What is the *fraction* of time that a transducer is transmitting? - RIGHT ANSWER -duty factor
What is the accuracy of reflector depth called? - RIGHT ANSWER -depth calibration
What is the accuracy of reflector depth position on an A, B, or M-mode display called? - RIGHT ANSWER -depth calibration
What is the actual time that an US machine is creating a pulse? - RIGHT ANSWER -pulse duration
What is the approximate attenuation coefficient of 1 MHz US in soft tissue? - RIGHT ANSWER -.5 dB/cm
What is the elimination of voltages that do NOT exceed a certain level? - RIGHT ANSWER -reject
What is the first thing you should do when entering a hospital room? - RIGHT ANSWER -check the patient's wristband for identification
What is the Fraunhofer zone? - RIGHT ANSWER -far zone
What is the frequency of a wave with a 1 msec period? - RIGHT ANSWER -1 KHz
What is the greatest advantage of one and one-half dimensional array transducers? - RIGHT ANSWER -improved slice thickness resolution
What is the location of the minimum cross-sectional area of the US beam called? - RIGHT ANSWER -focus
What is the meaning of a thermal index = 3? - RIGHT ANSWER -tissue temperature may rise 3 degrees celsius
What is the minimum value of the duty factor? - RIGHT ANSWER -0%
What is the most common cause of localized vertical non-uniformity in a real-time B-mode image? - RIGHT ANSWER -defective transducer elements
What is the most typical Doppler shift measured clinically? - RIGHT ANSWER -2 kHz
What is the name of the receiver function that corrects for attenuation? - RIGHT ANSWER -compensation
What is the Nyquist limit if a pulsed-wave doppler instrument has a pulse repetition frequency of 16 kHz? - RIGHT ANSWER -8 kHz
What is the propagation speed determined by? - RIGHT ANSWER -density and elasticity
What is the range of frequencies created by medical diagnostic Doppler transducers? - RIGHT ANSWER -2-10 MHz
What is the significance of gray scale variation in the spectrum of a pulse wave Doppler display? - RIGHT ANSWER -it represents the number of RBC's creating the reflection
What is the standard for communication between medical imaging computers? - RIGHT ANSWER -DICOM
What is the wavelength of a 3 MHz sound pulse in soft tissue? - RIGHT ANSWER -.49 mm
What may be a unit of amplitude? - RIGHT ANSWER -cm
What measures the output of a transducer? - RIGHT ANSWER -hydrophone
What will increase signal-to-noise ratio? - RIGHT ANSWER -increasing output power
What will NOT improve temporal resolution? - RIGHT ANSWER -increased line density
When a fluid is flowing at a slow and constant rate within a long tube having a uniform diameter, which is the nature of the flow? - RIGHT ANSWER -laminar
When may a patient revoke their consent to be treated? - RIGHT ANSWER -at any time
When the focal point in a B-mode image is electronically changed, which type of focusing is in use? - RIGHT ANSWER -dynamic
When the US machine displays only strong reflecting objects and nothing else, the sonographer should: - RIGHT ANSWER -increase output power
Where are harmonics created? - RIGHT ANSWER -in the tissues
Where does coded excitation originate? - RIGHT ANSWER -the pulser
Which action can improve resolution of a displayed image? - RIGHT ANSWER -increasing the transducer frequency
Which angle of insonation to blood flow produces the maximum doppler shift: 180, 150, 90, 45 degrees - RIGHT ANSWER -180 (same as zero)
Which best describes annular array transducers? - RIGHT ANSWER -excellent lateral resolution, diminished temporal resolution, sector shaped
Which component of an ultrasound system exposes the patient to the greatest risk? - RIGHT ANSWER -transducer
Which control allows adjustment for attenuation at specific depths? - RIGHT ANSWER -TGC
Which factor determines the displayed depth of a reflector? - RIGHT ANSWER -PRF
Which factor reduces the likelihood of aliasing in pulsed wave spectral doppler? - RIGHT ANSWER -increased nyquist limit
Which has the *greatest* amount of attenuation? - RIGHT ANSWER -bone
Which imaging mode requires two elements mounted side by side? - RIGHT ANSWER -CW doppler
Which is most likely to create cavitation? - RIGHT ANSWER -high peak rarefaction pressure and low frequency
Which is not a part of the transducer? - RIGHT ANSWER -A/D converter
Which is not an acoustic variable? - RIGHT ANSWER -intensity
Which of the following best describes a linear sequential array transducer? - RIGHT ANSWER -dynamic receive focusing, rectangular shaped
Which of the following best describes a sound beam? - RIGHT ANSWER -a collection of mechanical vibrations in the form of compressions and rarefactions
Which of the following best describes a vector array transducer? - RIGHT ANSWER -electronic steering and focusing, trapezoidal shape
Which of the following best describes demodulation of electronic signals? - RIGHT ANSWER -changing the shape or form of the signal
Which of the following best describes the *empirical* approach to the study of bioeffects? - RIGHT ANSWER -exposure-response
Which of the following best describes the *mechanistic* approach to the study of bioeffects? - RIGHT ANSWER -cause-effect
Which of the following best describes the features of a sound wave and medium that determine the speed of sound in the medium? - RIGHT ANSWER -density and stiffness of the medium
Which of the following choices best describes an ideal epidemiological bioeffects study? - RIGHT ANSWER -randomized and prospective
Which of the following choices best describes axial resolution artifact? - RIGHT ANSWER -two reflectors positioned in front of the other appears as one reflection
Which of the following choices determines the signal amplitude in the transducer of an ultrasound system? - RIGHT ANSWER -reflected intensity of the sound beam
Which of the following choices determines the signal amplitude in the transducer of an ultrasound? - RIGHT ANSWER -reflected intensity
Which of the following clinical modalities has the lowest output intensity? - RIGHT ANSWER -gray scale imaging
Which of the following describes the situation where the minimum amount of attenuation will occur? - RIGHT ANSWER -4 cm depth and 6 MHz frequency
Which of the following forms of cavitation are most likely to produce micro-streaming in the intracellular fluid and shear stresses? - RIGHT ANSWER -stable cavitation
Which of the following forms of focusing is achieved during reception? - RIGHT ANSWER -dynamic aperture
Which of the following functions of a receiver is used to create similar brightness reflections from identical reflectors positioned at different depths in the body? - RIGHT ANSWER -compensation
Which of the following has a propagation speed closes to soft tissue? - RIGHT ANSWER -muscle
Which of the following has the greatest influence on temporal resolution? - RIGHT ANSWER -propagation speed and image depth
Which of the following has the greatest output intensity? - RIGHT ANSWER -pulsed doppler
Which of the following improves signal-to-noise ratio? - RIGHT ANSWER -frame averaging
Which of the following is a benefit of using a contrast agent? - RIGHT ANSWER -increased amplitude of the returning doppler signals
Which of the following is associated with cavitation? - RIGHT ANSWER -mechanical index
Which of the following is associated with power Doppler? - RIGHT ANSWER -poor temporal resolution and flash artifact
Which of the following is determined by the source and the medium? - RIGHT ANSWER -range resolution
Which of the following is most directly related to temporal resolution? - RIGHT ANSWER -frame rate
Which of the following is not a beam steering technique for a mechanical scanhead? - RIGHT ANSWER -phased beam
Which of the following is NOT used to measure the output of ultrasound systems? - RIGHT ANSWER -calibrated absorpterometer
Which of the following is proportional to the square of the amplitude? - RIGHT ANSWER -power
Which of the following is the best lateral resolution? - RIGHT ANSWER -.06 cm
Which of the following is the cause of slice thickness artifact? - RIGHT ANSWER -beam dimension greater in size than the reflector
Which of the following is the correct order of receiver operation? - RIGHT ANSWER -amplification, demodulation, reject
Which of the following is the spectral analysis technique of choice for PW and CW Doppler? - RIGHT ANSWER -FFT
Which of the following lists is in decreasing order? kilo, deci, micro, hecto, milli, nano - RIGHT ANSWER -kilo, hecto, deci, milli, micro, nano
Which of the following situations should result in a cessation of the use of diagnostic ultrasound? - RIGHT ANSWER -the benefits are out-weighted by the risks
Which of the following sound waves is ultrasonic and least useful? - RIGHT ANSWER -30 KHz
Which of the following statements about Doppler is most correct? - RIGHT ANSWER -non-directional Doppler uses non-coherent processing, while bi-directional Doppler uses phase-quadrature processing
Which of the following statements related to venous flow is most correct? - RIGHT ANSWER -during inspiration, venous flow in the arms increases & during expiration, venous flow in the legs increases
Which of the following steps is performed just prior to storage of image data in a digital scan converter? - RIGHT ANSWER -A-to-D conversion
Which of the following tissues will most likely create Rayleigh scattering? - RIGHT ANSWER -red blood cells
Which of the following transducers will create a sound beam with the least amount of divergence? - RIGHT ANSWER -high frequency, large diameter
Which of the following transducers will create sound with the lowest frequency? - RIGHT ANSWER -thick crystal, PZT in which sound travels slowly
Which of the following will have the lowest SPTA intensity at a depth of one near zone? - RIGHT ANSWER -4 MHz, 12 mm diameter
Which of the following will reduce ghosting artifact? - RIGHT ANSWER -wall filter
Which of the following will result in the greatest number of shades of gray in a digital image display? - RIGHT ANSWER -large pixels with many bits/pixel
Which of the following will result in the highest mechanical index? - RIGHT ANSWER -low frequency and high signal amplitude
Which of these biologic tissues will sound waves propagate most quickly? - RIGHT ANSWER -tendon
Which of these boundaries will create the strongest reflection? - RIGHT ANSWER -air-soft tissue
Which of these boundaries will create the weakest reflection? - RIGHT ANSWER -soft tissue-blood
Which of these cannot be performed by the sonographer? - RIGHT ANSWER -clean pixels
Which of these factors has the greatest influence on PRP? - RIGHT ANSWER -display depth
Which of these intensities is the smallest? SATP, SPTP, SATA - RIGHT ANSWER -SATA
Which of these statements is true of a 25 MHz sound pulse? - RIGHT ANSWER -highly attenuating, short SPL
Which of these terms does NOT belong with the others?
high duty factor, shallow imaging, short pulse duration, short PRP, high PRF - RIGHT ANSWER -short pulse duration
Which of these terms does NOT belong with the others? gray scale, few choices, uncompressed signal, wide dynamic range, low contrast - RIGHT ANSWER -few choices
Which of these terms means "many gray shades"? - RIGHT ANSWER -wide dynamic range
Which one of the following sets of properties of a test phantom is most relevant when assessing depth calibration accuracy? - RIGHT ANSWER -reflector spacing and propagation speed
Which receiver function cannot be adjusted by the sonographer? - RIGHT ANSWER -demodulation
Which setting of an ultrasound system gives the sonographer the ability to display both strong and weak echoes? - RIGHT ANSWER -dynamic range
Which technique increases the sharpness of high contrast boundaries? - RIGHT ANSWER -edge enhancement
Which technique results in a 3D image? - RIGHT ANSWER -surface rendering
Which technique utilizes modulation of the transmitted pulse to improve signal-to-noise ratio? - RIGHT ANSWER -coded excitation
Which transducer would be best to image superficial structures? - RIGHT ANSWER -small diameter, high frequency
Which type of artifact appears most commonly with highly reflective objects? - RIGHT ANSWER -shadowing
Which type of interaction between sound and tissue is most affected by the thermal index? - RIGHT ANSWER -absorption
Which type of transducer will have the shallowest focus? - RIGHT ANSWER -low frequency, small diameter
Which variable decreased when the number of acoustic lines per frame is increased without changing the maximum depth? - RIGHT ANSWER -frame rate
While propagating, a sound wave is refracted. Finally, it returns to the transducer. The system processes the sound wave as if it has: - RIGHT ANSWER -traveled in a straight line, reflected, and returned
While scanning in a water tank, what location in a sound beam has the highest intensity? - RIGHT ANSWER -focal zone
With an unfocused transducer, what is the area from the face of the transducer to where the beam diverges called? - RIGHT ANSWER -near field
With good images on the display but not on the printer, one should: - RIGHT ANSWER -adjust the recorder
With normal incidence, what factors affect refraction? - RIGHT ANSWER -none of the above
With phased array transducers, tiny time delays are used to steer the sound beams in different directions. Which component of an ultrasound system creates these time delays? - RIGHT ANSWER -beam former
With the presence of __________, we see more boundaries than there are reflecting surfaces - RIGHT ANSWER -reverberations
You are performing a quality assurance study on an ultrasound system using a tissue equivalent phantom. You make an adjustment using a knob on the system console that changes the appearance of reflector brightness from fully bright to barely visible. What are you evaluating? - RIGHT ANSWER -system sensitivity
You enter a room only to find the patient does not speak English. Which of the following is best to obtain informed consent? - RIGHT ANSWER -call for a hospital interpreter
Vista previa del contenido
SPI PRACTICE QUESTIONS 2024/2025
_________ is a redirection of US in many directions as a result of a *rough* boundary between
two media - RIGHT ANSWER -scattering
___________ resolution is determined by SPL - RIGHT ANSWER -radial or axial
______________ is the distance covered by one cycle - RIGHT ANSWER -wavelength
______________ is the time to complete one cycle - RIGHT ANSWER -period
A circular hyperechoic structure is located at a depth of 5 cm. The structure has abnormally low
attenuation and the speed in the mass is 1,400 m/s. Which of the following is most likely to
appear beneath the structure? - RIGHT ANSWER -reflectors beneath the structure appear
abnormally deep in the image
A Doppler exam is performed with a 5 MHz probe and a PRF of 15 kHz. Which of these Doppler
shifts will create aliasing? - RIGHT ANSWER -14 kHz
A hydrophone is used to measure: - RIGHT ANSWER -intensity (or frequency)
A pulse is emitted by a transducer and is traveling in soft tissue. The go-return-time, or
time-of-flight, of a sound pulse is 130 microseconds. What is the reflector depth? - RIGHT
ANSWER -10 cm
A quality assurance program: - RIGHT ANSWER -is needed by all labs
A shadow appears on an ultrasound scan. Which of the following statements is true about the
shadow artifact? - RIGHT ANSWER -it appears as hypoechoic
,A shadow appears on an ultrasound scan. Which of the following statements is true about the
structure immediately above the shadowing artifact? - RIGHT ANSWER -attenuation properties
are too high
A single element transducer is focused: - RIGHT ANSWER -by the manufacturer
A sound beam is created by a single, disc-shaped crystal. Which of these beams has the
shallowest focus? - RIGHT ANSWER -2 MHz, 9 mm diameter
A sound wave is traveling in the body and propagating from one form of soft tissue to another.
For example, sound is moving from muscle to blood. What percentage of the sound wave is most
likely to be reflected at the muscle-blood boundary? - RIGHT ANSWER -.5%
A sound wave with a frequency of 5 MHz is travelling in soft tissue. What is the wavelength? -
RIGHT ANSWER -.3 mm
A sound wave's frequency is 10 MHz. The wave is traveling in soft tissue. What is its wavelength?
- RIGHT ANSWER -.15 mm
A system which is not working well now does not display clear images in the far field. This may
be a problem with: - RIGHT ANSWER -transducer, receiver, monitor, or scan converter
A transducer converts __________ from one form to another - RIGHT ANSWER -energy
A transducer with more than one active element is called: - RIGHT ANSWER -array
According to AIUM, there have been no confirmed bioeffects below intensities of ______ watts
per square centimeter SPTA - RIGHT ANSWER -.01
,Aliasing is an artifact observed when performing Doppler ultrasonography. Which of the
following signals is most likely to alias? - RIGHT ANSWER -7 MHz with a PRF of 3 kHz
All of the following are limitations of analog scan converts EXCEPT: - RIGHT ANSWER -low spatial
resolution
All of the following are primarily used to change the overall brightness of an entire image EXCEPT:
- RIGHT ANSWER -compensation
All of the following are true EXCEPT: - RIGHT ANSWER -two waves with identical frequencies
must interfere constructively
All of the following are true EXCEPT: - RIGHT ANSWER -two waves with identical frequencies
must interfere constructively
All of the following are true of normal cavitation EXCEPT: - RIGHT ANSWER -harmful bioeffects
are commonly observed
All of the following are true regarding temporal resolution EXCEPT: - RIGHT ANSWER -decreased
temporal resolution is associated with shorter pulse duration
All of the following artifacts place the erroneous reflection below the true reflector EXCEPT: -
RIGHT ANSWER -side lobe
All of the following describe dynamic range EXCEPT: logarithmic, comparison, relative, absolute,
ratio - RIGHT ANSWER -absolute
, All of the following regarding the hemodynamics of a stenosis are true EXCEPT: - RIGHT ANSWER
-pressure is lowest downstream
All of the following result from the use of backing material in a transducer EXCEPT: - RIGHT
ANSWER -high sensitivity
All of the following statements regarding in-vitro bioeffects are true EXCEPT: - RIGHT ANSWER
-in-vitro bioeffects conclusions are clinically relevant
All of the following will improve temporal resolution EXCEPT: - RIGHT ANSWER -increased line
density
An image is too dark. Which technique is best to correct this problem? - RIGHT ANSWER
-increase receiver gain
An individual is standing. The blood pressure generated by the heart is 120 mmHg. What is the
hydrostatic pressure at the ankle? - RIGHT ANSWER -100 mmHg
An ultrasound system is imaging to a depth of 10 cm. Which of the following will be different on
a single image at depths of 2 cm, 5 cm, and 10 cm? - RIGHT ANSWER -lateral resolution
An ultrasound system is operating in the harmonic mode. The image on the display of the
ultrasound system was created using sound reflections with a frequency of 6 MHz. What is the
frequency of sound emitted by the system's transducer? - RIGHT ANSWER -3 MHz
An ultrasound system is set at 0 dB and is transmitting at full intensity. What is the output power
when the system is transmitting at 10% of full intensity? - RIGHT ANSWER --10 dB
An ultrasound system is set at 0 dB and is transmitting at full intensity. What is the output power
when the system is transmitting at 50% of full intensity? - RIGHT ANSWER --3 dB