ANSWER All of the above
You have just received an Aribex NOMAD Pro in the rugged carrying case. Which of the
following is NOT considered a preliminary check? - ANSWER Determine if both handsets are
fully charged
If the unit is powered on, then set down and not touched, it will power itself off after - ANSWER
3 mins
The Aribex NOMAD Pro allows the operator to customize settings according to individual
preferences - ANSWER True
For added security, when finished for the day, remove the handset and secure it in a separate
area from main unit - ANSWER True
The best way to routinely clean the Aribex NOMAD Pro is to - ANSWER Follow manufacturers
suggestions or local infection control protocols
"SYSTEM FAILURE" is displayed on the screen. What is the appropriate user-level corrective
action in this situation? - ANSWER The unit detected a potential internal problem during the
power up self-test. Turn the unit off, and then back on. If there is no issue detected, continue w/
normal usage
While preparing the patient for an X-ray, you determine that it will not be possible to place the
backscatter shield and collimator tube 90 degrees to the image receptor. This may place you
outside the protection zone. What safety measure should you take? - ANSWER Use of a lead
apron, in accordance w/ local radiation safety protocols, is recommended