Who developed the international 10/20 placement? Ans✓✓-- Herbert Jasper
- up to 256 electrode sites exist (excludes reference electrodes)
*in 1949
What are the recommended parameters for sleep? (8) Ans✓✓-1. EEG
2. EKG
3. EOG
4. EMG
5. O2 monitoring
6. Belts
7. Airflow
8. Body position
What is the recommended EEG montage for sleep? Ans✓✓-F4 - M1
C4 - M1
O2 - M1
M1 - M2
What are the recommended backup leads in an EEG montage for sleep? Ans✓✓-
F3 - M2
C3 - M2
, O1- M2
What areas need to be recorded from during sleep? Ans✓✓-frontal, central,
occipital
What are the recommended EEG filter settings? (SR, LFF, HFF) Ans✓✓-SR- 500Hz
(desirable)
LFF- 0.3Hz
HFF- 35Hz
*SR range 500Hz-200Hz minimal
What are the recommended EOG filter settings? (SR, LFF, HFF) Ans✓✓-SR- 500Hz
(desirable)
LFF- 0.3Hz
HFF- 35Hz
*SR range 500Hz-200Hz minimal
What are the recommended EKG filter settings? (SR, LFF, HFF) Ans✓✓-SR- 500Hz
(desirable)
LFF- 0.3Hz
HFF- 70Hz
*SR range 500Hz-200Hz minimal
*referenced to each other