CCHT WATER questions and answers with solutions 2025
CHLORINE TESTING - ANSWER REVERSE OSMOSIS (RO) MACHINES RUNS 15 MINUTES BEFORE PERFORMING ANY TOTAL CHLORINE TESTING 100 ML - ANSWER THE CORRECT AMOUNT OF WATER TO COLLECT FOR TOTAL CHLORINE TESTING RPC ultra low total chlorine strips - ANSWER STRIPS USED FOR CHLORINE TESTING HOW LONG ARE RPC STRIPS - ANSWER IMMERSED IN SAMPLE WATER FOR 60 SECONDS WHEN SHOULD CHLORINE BE TESTED - ANSWER PRIOR TO TREATMENTS IN THE MORNING AND EVERY 3-4 HOURS WHILE PATIENTS ARE DIALYZING RPC ULTRA -LOW TOTAL CHLORINE STRIPS TEST - ANSWER CHLORINE IN THE WATER AFTER CARBON TANK AND POST RO 0.00 to 0.09 ppm - ANSWER total chlorine levels that are patient safe AND DOCUMENT IN "TMS" OR PAPER LOG FIRST SAMPLE - ANSWER SHOULD BE TAKEN FROM POST CARBON TANK 1 (WORKER ) SECOND SAMPLE - ANSWER POST RO IF CHLORINE IS NOTED POST RO - ANSWER DO NOT INITIATE ANY PATIENT TREATMENT, AND DOCUMENT PER FACILITY POLICY AND PROCEDURE CHLORINE BREAKTHROUGH - ANSWER OCCURS POST CARBON TANK 1 (WORKER), IMMEDIATE ADDITIONAL TESTING SHOULD TAKE PLACE AFTER CARBON TANK 2 (POLISHER) CARBON TANK 2 (POLISHER) - ANSWER IF