Escrito por estudiantes que aprobaron Inmediatamente disponible después del pago Leer en línea o como PDF ¿Documento equivocado? Cámbialo gratis 4,6 TrustPilot
logo-home
Document preview thumbnail
Vista previa 2 fuera de 14 páginas
Examen

CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.

Document preview thumbnail
Vista previa 2 fuera de 14 páginas

CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR SUCCESS.

Vista previa del contenido

CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST
UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR
SUCCESS.
Sterilizer Anatomy - CORRECT ANSWERS Drain line of the most important parts to check, is the
coldest part of the sterilizer because this is where the air & steam exit. Where the thermometer is
located and where the PCD/Bowie dick test pack is placed for testing because it is the location of
greatest challenge in the steam sterilizer.

door gasket the second most important to check

Sterilizer jacket is where steam is kept warm and surrounds the internal chamber.

Internal chamber where the steam purged into the chamber should be 97-100% dry saturated steam
w/3% entrained water.



simultaneous wrapping - CORRECT ANSWERS Use of two thicknesses of a wrapping material at the
same time.



Sequential wrapping - CORRECT ANSWERS wrapping one wrap after another.



Rigid Scopes - CORRECT ANSWERS Arthroscope, Hysteroscope, Resectoscope, Laparoscope,
Cystoscope, Nephroscope, Otoscope



Soft cloth and detergent - CORRECT ANSWERS Should be used to clean fiberoptic light cables



Woven Textiles - CORRECT ANSWERS Wrap is made of 100% cotton, reusable, & have 30- day
expiration date. they are also not a good barrier.



woven textiles can be used in - CORRECT ANSWERS Steam, EO/ ETO



Packing Materials: Non Woven - CORRECT ANSWERS wraps is made of cellulose fibers-plastic
polymers-paper pulp, are single use and provide an excellent barrier.



Non-woven can be used in - CORRECT ANSWERS Steam, EO/ETO, Low Temperature Gas Plasma



Packing Materials: Tyvek - CORRECT ANSWERS pouches are made of all plastic and should never be
put in trays.



Tyvek can be used in - CORRECT ANSWERS EO/ETO, Low Temperature Gas Plasma,Ozone

, CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST
UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR
SUCCESS.

Tyvek can not be used in - CORRECT ANSWERS Steam



BI placement: Steam - CORRECT ANSWERS over the drain



BI placement: ETO - CORRECT ANSWERS center of load



BI placement: LTGP - CORRECT ANSWERS back bottom shelf on top of tray/pack



BI placement: LTHP - CORRECT ANSWERS center top shelf



BI placement: Ozone - CORRECT ANSWERS upper shelf



BI placement: Dry Heat - CORRECT ANSWERS fullest densest load



Sterile field wrappers should be - CORRECT ANSWERS 6 inches over the edge of the table in the OR.



set/pack weights Trays - CORRECT ANSWERS 25 lbs.



set/pack weights Basins - CORRECT ANSWERS 7 lbs.



set/pack weights Textiles - CORRECT ANSWERS 12 lbs.



immediate-use steam sterilization (IUSS) - CORRECT ANSWERS Used for one of a kind instruments
or instrumentation that is needed on an emergency basis only if the IFU states which is the first thing
that must be checked prior to IUSS of any instrumentation.



Superheated Steam - CORRECT ANSWERS Heated steam missing moisture. Superheating is heat
added to dry steam without water.

Información del documento

Subido en
4 de noviembre de 2024
Número de páginas
14
Escrito en
2024/2025
Tipo
Examen
Contiene
Desconocido
$21.99

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
Creativepdf
3.8
(5)
Vendido
38
Seguidores
28
Artículos
2597
Última venta
5 meses hace


Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes