UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR
SUCCESS.
Sterilizer Anatomy - CORRECT ANSWERS Drain line of the most important parts to check, is the
coldest part of the sterilizer because this is where the air & steam exit. Where the thermometer is
located and where the PCD/Bowie dick test pack is placed for testing because it is the location of
greatest challenge in the steam sterilizer.
door gasket the second most important to check
Sterilizer jacket is where steam is kept warm and surrounds the internal chamber.
Internal chamber where the steam purged into the chamber should be 97-100% dry saturated steam
w/3% entrained water.
simultaneous wrapping - CORRECT ANSWERS Use of two thicknesses of a wrapping material at the
same time.
Sequential wrapping - CORRECT ANSWERS wrapping one wrap after another.
Rigid Scopes - CORRECT ANSWERS Arthroscope, Hysteroscope, Resectoscope, Laparoscope,
Cystoscope, Nephroscope, Otoscope
Soft cloth and detergent - CORRECT ANSWERS Should be used to clean fiberoptic light cables
Woven Textiles - CORRECT ANSWERS Wrap is made of 100% cotton, reusable, & have 30- day
expiration date. they are also not a good barrier.
woven textiles can be used in - CORRECT ANSWERS Steam, EO/ ETO
Packing Materials: Non Woven - CORRECT ANSWERS wraps is made of cellulose fibers-plastic
polymers-paper pulp, are single use and provide an excellent barrier.
Non-woven can be used in - CORRECT ANSWERS Steam, EO/ETO, Low Temperature Gas Plasma
Packing Materials: Tyvek - CORRECT ANSWERS pouches are made of all plastic and should never be
put in trays.
Tyvek can be used in - CORRECT ANSWERS EO/ETO, Low Temperature Gas Plasma,Ozone
, CBSPD EXAM QUESTIONS AND CORRECT ANSWERS LATEST
UPDATE 2023/2024 BEST EXAM SOLUTION GRADED A+ FOR
SUCCESS.
Tyvek can not be used in - CORRECT ANSWERS Steam
BI placement: Steam - CORRECT ANSWERS over the drain
BI placement: ETO - CORRECT ANSWERS center of load
BI placement: LTGP - CORRECT ANSWERS back bottom shelf on top of tray/pack
BI placement: LTHP - CORRECT ANSWERS center top shelf
BI placement: Ozone - CORRECT ANSWERS upper shelf
BI placement: Dry Heat - CORRECT ANSWERS fullest densest load
Sterile field wrappers should be - CORRECT ANSWERS 6 inches over the edge of the table in the OR.
set/pack weights Trays - CORRECT ANSWERS 25 lbs.
set/pack weights Basins - CORRECT ANSWERS 7 lbs.
set/pack weights Textiles - CORRECT ANSWERS 12 lbs.
immediate-use steam sterilization (IUSS) - CORRECT ANSWERS Used for one of a kind instruments
or instrumentation that is needed on an emergency basis only if the IFU states which is the first thing
that must be checked prior to IUSS of any instrumentation.
Superheated Steam - CORRECT ANSWERS Heated steam missing moisture. Superheating is heat
added to dry steam without water.