Ohm's law equation - ANSWER-V=IR EKG monitored patient should be __________, monitor chassis should be ___________ (nonconductive). Why - ANSWER-ungrounded, grounded. Grounded patient would be able to conduct leakage currents. we dont want this lb to kg - ANSWER-2.2 lb = 1 kg F to C - ANSWER-C=5/9(F-32) C to F - ANSWER-F=9/5C+32 3 things that change automaticity of automatic cells - ANSWER-1. slope of phase 4 2. Change of threshold potential 3. Change of resting membrane potential a systole is represented by what portion of the ECG - ANSWER-QT interval What is main ion transfer during systole - ANSWER-K+ seeps out, Ca2+ enters At rest, what is intracellular/extracellular environment - ANSWER-inside negative, outside positive What phase of AP is resting membrane potential - ANSWER-phase 4 Nodal depolarization occurs via what channel - ANSWER-Ca2+ slow channel Cardiac muscle cell depol occurs via what channel - ANSWER-Na+ fast channel During which phases is cell refractory - ANSWER-1,2,3 TDP caused by what mechanism usually - ANSWER-EAD's (triggered activity) Variance - ANSWER-Take each number in sample, subtract each number from mean, square each difference, add all, divide sum by number in sample -1 Standard deviation - ANSWER-square root of variance P value - ANSWER-Probability of certainty / smaller p value means the more likely the result could not occur by chance survival curve showing % of patients surviving treatment over time - ANSWER-Kaplan - Meier curves when measuring EGM's, earliest atrial activation usually from _____ and earliest ventricular usually from ______ - ANSWER-P wave / QRS Morady maneuver is used to differentiate - ANSWER-differentiate AT from AVNRT/AVRT Most common arrhythmia - ANSWER-AF Most common SVT - ANSWER-AVNRT What type of single use devices can be resterilized and why? - ANSWER-Diagnostic EP electrodes because lumen and technically not a catheter because no lumen Fr to mm - ANSWER-1 Fr = .33 mm Tip inner diameter standard diagnostic cardiac catheters and why - ANSWER-.038 inch / because it is supposed to be used with .035-.038 inch guide wires What is 2 rules for CMC's like spiral or lasso -

Mostrar más Leer menos
Institución
IBHRE CEPS
Grado
IBHRE CEPS










Ups! No podemos cargar tu documento ahora. Inténtalo de nuevo o contacta con soporte.

Escuela, estudio y materia

Institución
IBHRE CEPS
Grado
IBHRE CEPS

Información del documento

Subido en
9 de octubre de 2024
Número de páginas
22
Escrito en
2024/2025
Tipo
Examen
Contiene
Preguntas y respuestas

Temas

Vista previa del contenido

IBHRE CEPS Exam Questions and
Answers(RATED A)


Ohm's law equation - ANSWER-V=IR

EKG monitored patient should be __________, monitor chassis should be
___________ (nonconductive). Why - ANSWER-ungrounded, grounded. Grounded
patient would be able to conduct leakage currents. we dont want this

lb to kg - ANSWER-2.2 lb = 1 kg

F to C - ANSWER-C=5/9(F-32)

C to F - ANSWER-F=9/5C+32

3 things that change automaticity of automatic cells - ANSWER-1. slope of phase 4
2. Change of threshold potential
3. Change of resting membrane potential

a systole is represented by what portion of the ECG - ANSWER-QT interval

What is main ion transfer during systole - ANSWER-K+ seeps out, Ca2+ enters

At rest, what is intracellular/extracellular environment - ANSWER-inside negative,
outside positive

What phase of AP is resting membrane potential - ANSWER-phase 4

Nodal depolarization occurs via what channel - ANSWER-Ca2+ slow channel

Cardiac muscle cell depol occurs via what channel - ANSWER-Na+ fast channel

During which phases is cell refractory - ANSWER-1,2,3

TDP caused by what mechanism usually - ANSWER-EAD's (triggered activity)

Variance - ANSWER-Take each number in sample, subtract each number from mean,
square each difference, add all, divide sum by number in sample -1

Standard deviation - ANSWER-square root of variance

P value - ANSWER-Probability of certainty / smaller p value means the more likely the
result could not occur by chance

,survival curve showing % of patients surviving treatment over time - ANSWER-Kaplan -
Meier curves

when measuring EGM's, earliest atrial activation usually from _____ and earliest
ventricular usually from ______ - ANSWER-P wave / QRS

Morady maneuver is used to differentiate - ANSWER-differentiate AT from
AVNRT/AVRT

Most common arrhythmia - ANSWER-AF

Most common SVT - ANSWER-AVNRT

What type of single use devices can be resterilized and why? - ANSWER-Diagnostic EP
electrodes because lumen and technically not a catheter because no lumen

Fr to mm - ANSWER-1 Fr = .33 mm

Tip inner diameter standard diagnostic cardiac catheters and why - ANSWER-.038
inch / because it is supposed to be used with .035-.038 inch guide wires

What is 2 rules for CMC's like spiral or lasso - ANSWER-1. Only approved for use in LA
due to chordae tendinea
2. Only rotate shaft clockwise

Stereotaxis catheters - ANSWER-Have magnets in tip for maneuverability in response
to changes in magnetic field

What system remotely steers guided catheters - ANSWER-Hansen robotic system

ACT for using array - ANSWER-ACT 250-300

bipolar cardiac electrodes record local _________ that occur during phase _______ -
ANSWER-depolarizations / phase 0

surface EKG leads usually filtered at what - ANSWER-.1 - 100 Hz

Normal EGM filter settings - ANSWER-30 - 300 Hz

What to do if T waves interfere with A/V and why - ANSWER-Increase the high pass
filter because T waves are lower frequency as well as far field

Unipolar filter setting - ANSWER-.05 - 300 Hz

How each is measured: Diagnostic catheter OD , Inflated balloon cath OD, Needle OD,
Guide wire - ANSWER-Diag: Fr / Balloon: mm / Needle: gauge / GW: inch (thousandth
of inches)

, Standard sizes of micropuncture introducer sets and their purpose - ANSWER-21
gauge needle / .018 inch wire. Purpose of micropuncture introducer set is for placement
of .035-.038 inch GW

Best needle to use for pericardial space via subxyphoid approach - ANSWER-Tuohy
needle

which type of guide wire is 260-300 cm long? - ANSWER-exchange guide wires (much
longer than diagnostic catheter)

advantage of bipolar leads - ANSWER-less prone to EMI noise

Unipolar leads more prone to - ANSWER-1. EMI / muscle artifact oversensing
2. pectoralis muscle stimulation (pocket stim)

Steroid tip leads reduce acute: - ANSWER-inflammation and stimulation threshold

Best numbers for ventricular lead - ANSWER-Threshold <1 V / Sensing >4 mV

programmed stimulation - ANSWER-Consists of incremental pacing and extrastimuli

Sharp, curved GW used to go transseptal - ANSWER-SafeSept Guidewire

Baylis NRG RF Transseptal needle has - ANSWER-Side holes for pressure and
contrast

Ultrasound transducers are composed of - ANSWER-Piezoelectric crystals

How do ultrasound transducers work - ANSWER-Transducers convert one form of
energy into another. Ultrasound transducers convert electric signals into ultrasonic
energy. The ultrasonic energy is transmitted into tissues, bounces back and then is
converted back to electric energy

doppler transducer uses what - ANSWER-single piezoelectric crystal that sends and
receives

In cardiovascular doppler, what is target off of which ultrasound waves are reflected
back - ANSWER-RBC's

ICE advantage over TEE - ANSWER-General anesthesia not needed for ICE

What is used to visualize shunts or distinguish right from left chamber - ANSWER-
Agitated saline

Where does conductive heating occur - ANSWER-2-5 mm beneath electrode
$11.99
Accede al documento completo:

100% de satisfacción garantizada
Inmediatamente disponible después del pago
Tanto en línea como en PDF
No estas atado a nada

Conoce al vendedor

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
IMORA West Virginia University
Seguir Necesitas iniciar sesión para seguir a otros usuarios o asignaturas
Vendido
195
Miembro desde
2 año
Número de seguidores
77
Documentos
5079
Última venta
1 día hace

4.4

69 reseñas

5
49
4
8
3
8
2
1
1
3

Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes