Fresenius Virtual Learning PCT New Hire Exam Review Questions And Answers
How long does the RO have to be running before we perform any water testing? - answer-15 minutes What does IFU stand for? - answer-instructions for use ***Instructions written by the manufacturer that are approved by the FDA. How long does the sample valve need to be flushed after the RO has been running for 15 minutes before a sample is collected? - answer-2 minutes What is the frequency for testing the Total Chlorine? - answer-Before the first patient of the day is put on AND every 4 hours after that. Name the test trip used to test Total Chlorine: - answer-RPC Ultra-Low Total Chlorine Strips What does TDS stand for? - answer-Total Dissolved Solids The RPC Ultra-Low Total Chlorine Strips are used to test the functionality of what tanks in the water room? - answer-GAC (Granular Activated Carbon) tanks Which tanks in the water room are responsible for removing Total Chlorine? - answer-GAC tanks (Granular Activated Carbon tanks)