DaVita PCT training exam questions with complete solutions
What is adequacy? - ANSWER--How well dialysis tx is working -measured by green post BUN lab - Kt/V goal is 1.2 or 65% What affects Kt/V? - ANSWER-- time of tx -frequency of tx -increasing BFR and DFR increase Kt/V Needle gauge sizes vs. BFR - ANSWER-- 17g=200-250 - 16g = 250-350 - 15g = 350-450 - 14g = 450 and ^ procedure for post BUN lab draw - ANSWER-turn off UF DFR to 300 decrease BFR to 100 ml/pm WAIT 15 SEC. draw from arterial line port What causes Kt/V to be falsely high? - ANSWER-if you take it before waiting the full 15 secs what causes kt/v to be falsely low? - ANSWER-if you wait longer than 15 sec to take it