Escrito por estudiantes que aprobaron Inmediatamente disponible después del pago Leer en línea o como PDF ¿Documento equivocado? Cámbialo gratis 4,6 TrustPilot
logo-home
Ensayo

Pearson BTEC Applied Science (Distinction) Unit 2: Scientific Procedures and Techniques Assignment 2A: Concentrate on keeping up your standards

Puntuación
-
Vendido
-
Páginas
28
Grado
A+
Subido en
21-07-2026
Escrito en
2022/2023

This assignment was completed as part of Unit 2: Scientific Procedures and Techniques for Pearson BTEC Applied Science. It was awarded a Distinction.

Institución
Grado

Vista previa del contenido

H.
BTEC L3 EXTENDED DIPLOMA IN APPLIED SCIENCES
Unit 2: Practical Scientific Procedures and
Techniques
Assignment 2A: Concentrate on keeping up your
standards

Laboratory equipment and its calibration

What is calibration?

Calibration is the scientific process of making comparisons between
measurements, by determining the relationship between values through
measurements and indications on a measuring instrument. This process is
performed in order to decrease any measurement uncertainty by guaranteeing
the accuracy of equipment.

Method to calibrate pH probe

Firstly, turn on the pH probe and thoroughly wash the pH probe with distilled
water in a sink.

Next, you must set up buffers to calibrate the pH probe. The first buffer solution
should be a neutral pH of 7, the second buffer solution should be close to the
estimated sample pH, e.g. pH 4 or 9po, depending on the whether the sample is
a base/acid and the third buffer solution should be one lower than the estimated
sample pH. For example, the buffer solutions I used for the calibration of the pH
probe was 6.86, 9.18 and 4.01. Once set up, leave the buffers to arrive at the
same temperature as the pH probe because pH changes with temperature,
therefore is temperature-dependent. Add the buffers into separate beakers in
order to calibrate the pH probe.

Following on from the last step, put the pH probe in the first beaker with the
neutral pH, and start the reading. To start the reading, push the
‘measure’/’calibrate’ button when the pH probe is placed in the beaker. Leave
the pH reading to settle before allowing it to set for around 1-2 minutes.

After attaining a stable reading, set the pH probe to the buffer’s pH by pushing
the ‘measure’ button again. Once this is done, wash the pH probe thoroughly
with distilled water, as done in the first step.

Put the pH probe in the next beaker and follow the same steps as the first
beaker, washing the pH probe when done.

Method to calibrate a burette/pipette

Firstly, to calibrate a pipette, ensure that it is completely clean and dry, as well
as empty, before calibration.

Following on to the next step, it involves using distilled water and you will need
to accurately measure the temperature of the distilled water using a
thermometer. To make sure the temperature obtained is accurate, leave the
thermometer in the water for at least 10 minutes.


1

, H.
BTEC L3 EXTENDED DIPLOMA IN APPLIED SCIENCES
Unit 2: Practical Scientific Procedures and
Techniques

Once the temperature is obtained, weigh a clean and dry beaker with a balance
(should be calibrated before). Then, use a pipette filler to extract the distilled
water until the 25ml graduation mark and transfer it to the beaker you have
weighed carefully. Next, record the mass of the beaker with the water using the
balance and repeat this step twice to obtain an average of the masses recorded.

Following on from the last step, you need to calculate the volume released by
the pipette using the formula volume = mass/density. The temperature obtained
from the distilled water is used to find the density of the water. For example, in
my calibration of the pipette, the temperature was 25 degrees Celsius which
meant the density was 0.996262. Once you have attained the density of the
water, you can use this with the average mass of all the readings you took on the
balance with the beaker of water. You will need to divide the average mass of
water by the density of the water and this value will be the volume released by
the pipette. For example, I calculated the average mass to be 26.0g and divided
this by the density of water – 26.0/0.996262 = 26.097 = 26.1 cm 3

Method to calibrate a balance

Firstly, place the balance on a flat surface and turn on the balance Figure 2 – Image of a
by pushing the ‘on/off’ button. balance being calibrated
with weights
Secondly, press and hold the ‘cal’ button on the balance in order to
calibrate the balance. When the balance display shows ‘cal 0’, let go of
the ’cal’ button. Then add 1-5 100g weight(s) onto the balance, and
wait for a reading on the balance display (as shown in Figure 2). If the
reading display shows the correct balance of the weight(s), e.g.2 x
100g weights should show 200g, press the ‘cal’ button again to confirm
the weight, allowing the balance to calibrate. If the weight shown on
the display is not correct, repeat the steps above until the weight
shown is correct.

Importance of calibrating equipment before undertaking an investigation

It is important to calibrate equipment beforehand as it determines the accuracy
and quality of results produced by equipment. As equipment is used over long
periods of time, the accuracy of results may shift especially when using
technologies, hence why calibration is necessary for equipment. To make sure
that the results we are producing are both accurate and reliable, we need to
continuously calibrate equipment before we carry out experiments.

The use of calibration is to reduce any uncertainties in measurements by
establishing the accuracy of equipment used. Therefore, calibration is important
as it allows us to control any errors/uncertainties that may be present within
measurements.

Examples of where calibration is important:




2

Escuela, estudio y materia

Nivel de Estudio
Editores
Tema
Curso

Información del documento

Subido en
21 de julio de 2026
Número de páginas
28
Escrito en
2022/2023
Tipo
ENSAYO
Profesor(es)
Desconocido
Grado
A+

Temas

$27.50
Accede al documento completo:

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Conoce al vendedor
Seller avatar
btecappliedsciencegradddd

Conoce al vendedor

Seller avatar
btecappliedsciencegradddd University College London
Seguir Necesitas iniciar sesión para seguir a otros usuarios o asignaturas
Vendido
-
Miembro desde
2 días
Número de seguidores
0
Documentos
33
Última venta
-

hiya thanks for viewing my page! selling my BTEC Applied Science assignments here :) on all my assignments i achieved a distinction and as a result was awarded D*D*D*! the optional units I did were 8, 9 , 10 , 12 , 14 and 21! note for unit 12 assignments 12B and 12D, I did the distinction criteria in a separate document- I have uploaded both documents with the distinction criteria and another with the merit and pass criteria! message me for any questions, advice or help!

Lee mas Leer menos
0.0

0 reseñas

5
0
4
0
3
0
2
0
1
0

Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes