Medical Gas Class

Category 1 medical -surgical vacuum sources shall, at minimum, consist of ______ or
more vacuum pumps - answer Two

Category 1 medical surgical vacuum sources shall consist of an automatic means too
prevent _____ through any off -cycle vacuum pump - answer Backflow

Piping between the vacuum pump and the source shutoff valve shall be permitted to be
stainless, galvanized, or black steel pipe - answer True

Vacuum receiver shall be capable of withstanding a gauge vacuum of ______ of
mercury - answer30"

The medical surgical vacuum receiver shall be made serviceable without shutting down
the system by _______ - answer Any method to ensure continuation service

When the medical surgical I'm an operation is incapable of adequately maintaining a
required vacuum, additional pump shall be _______ activated - answer automatically

The medical surgical vacuum exhaust shall be turned down and screened. The
screened shall be fabricated using what type of material? - answer Any non-corroding
material

Multiple vacuum pumps shall be permitted to be join together to come one common
exhaust provided the common exhaust is size to ______. - answer Minimize back
pressure

Vacuum exhaust from multiple pumps join together to one common exhaust shall be
permitted provided each pump can be isolated by a _____. - answerBlind flange,
manual valve, check valve

Category 1 medical-surgical vacuum systems shall activate which alarm when the
backup or lag pump is running? - answerLocal

A level 1 dedicated WAGD source shall consist of ______ or more producers, each
shall be sufficient to serve the peak calculated demand. - answerTwo

What source shall not be used to power the venturi of a WAGD system? -
answerMedical air

WAGD vacuum producers designed for operation at vacuums below ____ Hg/V shall be
used only for WAGD service and not employed for other services. - answer5"

, A WAGD source system shall activate the ____ alarm when the backup or lag producer
is running. - answerLocal

Additional WAGD producers shall be ________ activated when the producer(s) in
operation is incapable of maintaining required vacuum. - answerAutomatically

Medical air supply systems shall be required to _____. - answerMeet the requirements
of medical air USP, have no detectable liquid hydrocarbons, have less than 25 ppm
gaseous hydrocarbons

Medical air sources shall be used only for _____. - answerHuman respiratory
applications

Medical air compressor systems shall have a(n) ______ means to prevent back flow
through all off-cycle compressors - answerAutomatic

True - answerAn existing system that is not in strict compliance with the provisions of
this standard shall be permitted to be continued in use as long as the authority having
jurisdiction has determined that such use does not constitute a distinct hazard to life

True - answerCylinder and cryogenic liquid container outlet connections shall be
designed as to render the attachment point unusable when removal of the outlet
connection is attempted

Medical air sources - answerWhich of the following systems shall not be permitted to be
located in an indoor enclosure

Engines, kitchens, and anesthizing locations - answerIndoor locations for oxygen,
nitrous oxide and mixtures of these gases shall not communicate with ______ locations.

130 degree F - answerCylinders in use and in storage shall be prevented from reaching
temperatures in excess of ______.

True - answerLocations for central supply systems shall have electrical devices
protected from physical damage.

True - answerVentilation requirements shall not be less than 50 cfm for mechanical
ventilation.

72 sq. in. - answerLouvered openings for natural ventilation shall have a minimum free
area of _____.

Balanced - answerFinal line pressure regulators, when used for bulk cryogenic liquid
systems shall be of a _______ design.

Información del documento

Subido en
15 de mayo de 2026
Número de páginas
14
Escrito en
2025/2026
Tipo
Examen
Contiene
Preguntas y respuestas
$15.99

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
julianah420
4.2
(157)
Vendido
712
Seguidores
328
Artículos
35903
Última venta
3 días hace



Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes