Medical Gas Assessment 1, 2 & 3
A ____ device is used for connecting the outlets of one or more gas cylinders to the
central piping system? - answer Manifold
A factory assembled product designed for anesthetics or convenience that contains
medical gas outlets or vacuum inlets, piping, or other related to medical gases is a ____
? - answer Manufactured assembly
A warning system within an area uses a ____
that provides continuous visible and audible
surveillance of category 1 and category 2 medical gas and vacuum systems? - answer
Area alarm
After coolers where required shall be provided with individual ____ ? - answer
Condensate traps
Carbon monoxide shall activate a local alarms when the CO level exceeds ____? -
answer10 ppm
Compressors that are monitored for gases hydrocarbons shall be monitored on a ____
basis - answerQuarterly
Filters shall be efficient to ____ and 99.7% HEPA or better per. DOE-STD-3020 -
answer0.3
Medical air shall be sampled at a ____ NPS valve sample port downstream of the final
line pressure regulator? - answer1/4 inch
The minimum clearance around the EOSC for a temporary sore is ____ feet? - answer3
feet
The monitor shall be capable of monitoring ____ percent oxygen concentration with one
percent accuracy? - answer99
Within a single patient room or critical care area to provide access for two or more
gases through a common delivery system, which would be used? - answerSurface
mounted medical gas rail system also known as MGR
Which of the following systems shall not be permitted to be located in an outdoor
enclosure? - answermedical air sources
, What type of connection should be used to connect the medical air compressors for a
vacuum pumps to their intake an outlet piping? - answerFlexible
What is not part of a typical category one medical gas vacuum system? - answerPatient
lead
What is part of a typical category one medical gas vacuum system? - answerAn inlet at
locations where patient suction could be, pressure, gauges, and operating controls, shut
off valves
What is a pneumatic terminous of a pipe gas system? Other than critical continuous
duty, non-flammable, medical gas life support type gases. - answerService outlet
What guide is for preparation of precautionary, labeling, and marketing of compressed
gas containers? - answerCGA C-7
Vacuum filtration shall be sized for ____ percent of the peak calculated demand while
one filter or filter bundle is isolated. - answer100
Vacuum filtration shall be at least ____ to allow one filter to be exchanged without
impairing the vacuum system? - answerDuplex
Louvered openings for natural ventilation shall have a minimum free area of ____? -
answer72 square inches
Medical air filters shall be located immediately upstream of the ____? - answerFinal line
regulators
Cylinders and use an in storage shall be prevented reaching temperature is an excess
of ____? - answer125 Fahrenheit
All pressure relief valve shall be made of brass, bronze, or ____? - answerStainless
steel
A category one dedicated WAGD source shall consist of ____ or more producers, each
shall be sufficient to serve the peak calculated demand - answer2
A variant on the cryogenic liquid container manifold shall be permitted to have three
headers. Such a variant shall have a minimum of ____ connections required for each of
the primary and secondary headers. - answer2
For each oxygen concentrator a local alarm to indicate low oxygen concentration when
a concentration lower than ____ percent is observed? - answer91
Medical air compressors shall be provided with a ____ restart function, such that the
compressors will be restarted after a power interruption? - answerAutomatic
A ____ device is used for connecting the outlets of one or more gas cylinders to the
central piping system? - answer Manifold
A factory assembled product designed for anesthetics or convenience that contains
medical gas outlets or vacuum inlets, piping, or other related to medical gases is a ____
? - answer Manufactured assembly
A warning system within an area uses a ____
that provides continuous visible and audible
surveillance of category 1 and category 2 medical gas and vacuum systems? - answer
Area alarm
After coolers where required shall be provided with individual ____ ? - answer
Condensate traps
Carbon monoxide shall activate a local alarms when the CO level exceeds ____? -
answer10 ppm
Compressors that are monitored for gases hydrocarbons shall be monitored on a ____
basis - answerQuarterly
Filters shall be efficient to ____ and 99.7% HEPA or better per. DOE-STD-3020 -
answer0.3
Medical air shall be sampled at a ____ NPS valve sample port downstream of the final
line pressure regulator? - answer1/4 inch
The minimum clearance around the EOSC for a temporary sore is ____ feet? - answer3
feet
The monitor shall be capable of monitoring ____ percent oxygen concentration with one
percent accuracy? - answer99
Within a single patient room or critical care area to provide access for two or more
gases through a common delivery system, which would be used? - answerSurface
mounted medical gas rail system also known as MGR
Which of the following systems shall not be permitted to be located in an outdoor
enclosure? - answermedical air sources
, What type of connection should be used to connect the medical air compressors for a
vacuum pumps to their intake an outlet piping? - answerFlexible
What is not part of a typical category one medical gas vacuum system? - answerPatient
lead
What is part of a typical category one medical gas vacuum system? - answerAn inlet at
locations where patient suction could be, pressure, gauges, and operating controls, shut
off valves
What is a pneumatic terminous of a pipe gas system? Other than critical continuous
duty, non-flammable, medical gas life support type gases. - answerService outlet
What guide is for preparation of precautionary, labeling, and marketing of compressed
gas containers? - answerCGA C-7
Vacuum filtration shall be sized for ____ percent of the peak calculated demand while
one filter or filter bundle is isolated. - answer100
Vacuum filtration shall be at least ____ to allow one filter to be exchanged without
impairing the vacuum system? - answerDuplex
Louvered openings for natural ventilation shall have a minimum free area of ____? -
answer72 square inches
Medical air filters shall be located immediately upstream of the ____? - answerFinal line
regulators
Cylinders and use an in storage shall be prevented reaching temperature is an excess
of ____? - answer125 Fahrenheit
All pressure relief valve shall be made of brass, bronze, or ____? - answerStainless
steel
A category one dedicated WAGD source shall consist of ____ or more producers, each
shall be sufficient to serve the peak calculated demand - answer2
A variant on the cryogenic liquid container manifold shall be permitted to have three
headers. Such a variant shall have a minimum of ____ connections required for each of
the primary and secondary headers. - answer2
For each oxygen concentrator a local alarm to indicate low oxygen concentration when
a concentration lower than ____ percent is observed? - answer91
Medical air compressors shall be provided with a ____ restart function, such that the
compressors will be restarted after a power interruption? - answerAutomatic