QUESTIONS WITH COMPLETE SOLU
The cSI cunit cof clength cis cthe c- cCORRECT cANS✔✔Meter
The cPascal, cthe cSI cunit cof cpressure, cis c- cCORRECT cANS✔✔A cderived cSI cunit
Which cSI cmultipliers chave cabbreviations cin ccapital cletters? c- cCORRECT cANS✔✔Those cwith cfa
The cspeed cof clight cis cconsidered cto cbe ca cconstant c- cCORRECT cANS✔✔When cused cfor cstan
The cmelting cpoint cof cGallium cis ca cfixed cpoint cin cthe cITS-90 ctemperature csystem. cThe cuncerta
ctemperature cis c- cCORRECT cANS✔✔Zero
A ccertificate cfor ca ctraceable ccalibration cmust cshow cevidence cto csupport c- cCORRECT cANS✔✔
cpath cis cneeded cfor can caudit
The crealization cof ca ctemperature cfixed cpoint cis c- cCORRECT cANS✔✔A cphysical cobject csuch ca
Traceability cis crequired cwhen coperating cunder cISO c9001 cand crelated cstandards c- cCORRECT c
ccalibration cpyramid cbelow cnational cmeasurement cstandards
The cACME ccompany's ccalibration cprocedure cfor ctheir cmodel cVU-1 csound clevel cmeter ccalls cfor
csound clevel ccalibrator cmodel cSLC-1. cA ctechnician chas ca crush cjob cto ccalibrate ca cVU-1, cbut cth
cfactory cin cEurope cfor crepair cand ccalibration. cHe/she cshould c- cCORRECT cANS✔✔Stop cwork c
cthe ccalibration ccannot cbe ccompleted cuntil cthe cproper cIM&TE cis cavailable
In corder cto ccalibrate ca cdigital cammeter, cit cis cnecessary cto chave c- cCORRECT cANS✔✔A csourc
The cbase cSI cunit cfor celectrical cquantities cis c- cCORRECT cANS✔✔Ampere
One cgram cof coxygen cgas cis creleased cinto ca cbox c10 ccm con ca cside. cWhat cis cthe cSI cdensity co
cwhen cequilibrium cis creached? c- cCORRECT cANS✔✔1.0 ckg/m^3
, The cSI cunit cfor ctime cis cbased con c- cCORRECT cANS✔✔An catomic cvibration
Which cof cthe cfollowing cnotations caccepted cfor cuse cwith cthe cSI cis cpreferred cfor cthe cvolume cof
cCORRECT cANS✔✔3.785 cL c(1 cU.S. cgallon)
Carl cSagan, con chis ctelevision cseries, cmight chave csaid c"Thank cyou cfor cspending cthese cbillions
cwith cme." cHow clong ca ctime cinterval cis c1800 cbillion cns? cNote c- cuse cthe cU.S. cdefinition cof cbill
cANS✔✔Half can chour
Gravitational ccorrections care calways crequired cfor c- cCORRECT cANS✔✔Measurements cin cwhich
cforce
Which cof cthe cfollowing ctemperature cmeasuring cdevices crequires ceither can cactual cor ca csimulat
cCORRECT cANS✔✔Thermocouple
Crash cdummies cused cto csimulate cthe ceffects cof cadults cand cchildren cduring ctesting cof ccar cand
cexample cof ca c- cCORRECT cANS✔✔Consensus cstandard
Electrical csensors cused cto cmeasure ctemperature calmost calways crequire c- cCORRECT cANS✔✔
ctemperature c- celectrical crelationship
A csubstitute cstandard chas creplaced cthe cusual creference cstandard cand cthe cterm c"limited ccalibr
cWhich cof cthe cfollowing cexplanations cwould cNOT cbe cexpected cas can cexplanation cof cthe climite
cANS✔✔The csubstitute cwas cused cfor ca climited ctime
The cmost ccommon cworking cstandard cfor cdimensional canalysis cis c- cCORRECT cANS✔✔The cga
Measurement cof chumidity ccan cbe caccomplished c- cCORRECT cANS✔✔With ca cmirror cwhose cte
A claboratory cthat cuses can cintrinsic cstandard cmay cnot csimply cuse cit cwithout cverification, cbut cm
cANS✔✔Participate cin cinterlaboratory ccomparisons cor cproficiency ctesting cfor cthat cparameter
The crelationship cbetween cthe cfreezing cpoint cof cwater cand cthe ctriple cpoint cof cwater cat cstandar
cCORRECT cANS✔✔The ctriple cpoint cis cwarmer
The cSI cunit cused cby ctorque cwrenches cis c- cCORRECT cANS✔✔N*m
What cratio cof cattenuator cwould cbe crequired cto creduce ca c1.0 cmW csignal cto ca c1.0 cfW csignal? c
c000 c000 c000 c:1
The cbest creason cthat cthe clocal cacceleration cof cgravity, clittle cg, cneeds cto cbe cknown cby ca clabo
cCORRECT cANS✔✔It ccan caffect cmany ccalibration cresults
When cusing csaturated csalt csolutions cto cgenerate c(source) chumidity, cit cis cimportant cto cknow cth
cANS✔✔Temperate
A cclient chas csubmitted ca c0 c- c250 cft-lb, c+- c5% cof cfull cscale, ctorque cwrench cto ca clab cfor ccalibr
cdirection. cThe clab's c0 c- c600 cft-lb, c0.25% cof creading, cclockwise cand ccounter-clockwise, celectro
cThe clab calso cowns ca cspring-and-needle ctorque cclockwise ccalibrator cthat cis cworking cand cis cin
c1000 cft-lb, c1% cof creading c2 cft-lb. cCan cthis ccalibrator cbe cused cinstead? c- cCORRECT cANS✔✔
cjob
Which cof cthe cfollowing cparameters ccan cbe cconverted cto crelative chumidity cusing ca ctable cor cca
cANS✔✔Dew cpoint