When sound hits a boundary between two media of the same propagation speed at
oblique incidence: - ANSWER no refraction will occur.
Refraction requires the oblique incidence of sound across a boundary between two
media with different propagation speeds.
As transducer frequency increases: - ANSWER axial and lateral resolution will increase
Blood flow volume changes in the body are mainly controlled by what two factors? -
ANSWER cardiac function and peripheral resistance
When referring to blood flow volume throughout the body, the best answer to the
question is cardiac function and vessel resistance.
Potential bioeffects of ultrasound can be minimized by which of the following? -
ANSWER reduce exam time
Reducing overall exam time will reduce patient exposure. Adhering to the ALARA
principle will also reduce exposure by limiting power output adjustments and using TGC
controls instead. The other choices have no effect on intensity.
Which of the following types of resolution varies with different depths on the image? -
ANSWER lateral resolution
Lateral resolution is improved at the focal zone and in the near field compared to the far
field.
Transmit Delay or Receive Focusing improves _____________________ resolution. -
ANSWER angular and azimuthal
Azimuthal and angular resolution refer to lateral resolution. Lateral resolution is
improved with a thinner beam. Focusing forms a thinner beam in the focal zone,
therefore focusing improves the lateral resolution.
Which of the following is NOT adjustable by the Sonographer? - ANSWER
Demodulation
Demodulation occurs as the signal is converted into positive voltages (rectification) and
smoothed across the peaks of the voltage pulses. This process assists in processing
the voltages for proper image display of the returned signals. This is a preset function of
the US system and not operator adjustable.
To obtain a longer near zone: - ANSWER change from a 3MHz to a 6MHz transducer.
, Higher frequency and larger element diameter probes provide the longest near zones.
The matching layer for a transducer is usually made of plastic and is approximately
________________ . - ANSWER 1/4 of the probe wavelength in thickness
The matching layer for a transducer is usually made of plastic and is approximately 1/4
of the probe wavelength in thickness.
Lateral resolution is best at what location in the beam? - ANSWER focal point
If the ultrasound image is too black and white, which of the following will add more
levels
of grey to the image? - ANSWER decrease rejection levels
Rejection is used to reduce the amount of low level echoes displayed on the image. If
the image is too black and white, reducing the rejection level will allow more low level
echoes to be displayed on the image. This will decrease the black and white
appearance and increase the levels of grey displayed.
_______________ is defined as the time it takes to complete one cycle. - ANSWER
Period
Period is defined as the time it takes to complete one cycle.
The ability to differentiate a hyperechoic hemangioma within the hypoechoic liver tissue
is due to what type of image resolution? - ANSWER contrast resolution
Contrast resolution refers to the ability to differentiate different intensities (shades of
black, white and grey) on the ultrasound image. Compression can be used to alter the
dynamic range of the image which affects the contrast. Lower compression = larger
dynamic range = improved contrast resolution = lower image contrast = more shades of
gray on the image
Harmonic imaging leads to : - ANSWER decreased image clutter
Tissue harmonic imaging reduces noise and clutter by improving signal to noise ratio
and resolution. As an ultrasound wave propagates through the media, a change occurs
in the shape and frequency of the transmitted signal. The change is due to the normal
resistance of tissue/particles to propagate sound energy. This resistance and the
resulting signal change is called a harmonic oscillation.
The input frequency doubles the output frequency in harmonic imaging. EX: transmit
frequency of 4.0 MHz. for maximum penetration with a return harmonic frequency of 8.0
MHz for improved resolution. The returning higher frequency signal has to only travel
one direction to the probe. The advantages of high frequency imaging and the one-way