Laproscopic instruments diameter and length ranges - ANS 2-10mm, 30-45cm
R
hopkins rod lens - ANS light has to travel back through the rod to capture the image.
decreasing light in the camera for... - ANS decreasing diameter, increasing scope angle (ie
U
5mm and 30 degree has less light than 10mm 0 degree)
When is zero degree scope most useful - ANS when working in a small area directly in line
LA
with the scope and ports, like the pelvis
how to check fiber optics light connection - ANS black dots= broken fibers
why does it fog up? - ANS temperature and humidity discrepancy between the OR and
body
C
tools for defogging - ANS FRED antifog (must dry before putting back in), put laparoscope
in hot water
O
methods to clean a smudged lens - ANS gently wipe on clean tissue (liver, uterus, bowel),
remove scope and clean with hot water and gauze
D
Insufflation gas type and reasoning - ANS CO2- readily available, inexpensive, non
combustable, warmed and humidified better
high flow insufflation - ANS 10 or more L per minute
preventing loss of pneumo with suctioning - ANS keep suction tip below the fluid level
most common light source - ANS 300W xenon lamp
Troubleshooting steps: gas preOP - ANS 1. check that co2 tank is full