C.S. Certification Exam, SPD Test, Sterile
Processing Study Material For Certification Exam
Which area is the first to accept responsibility for all supplies in the facility?
Correct Ans - Receiving

Which type of inventory distribution system requires the use of a pre-
determined inventory list? Correct Ans - Par Level

Just-in-time distribution systems are also known as Correct Ans -
Stockless Distribution

Ordering the optimum quantity of an item based on historical data is known
as Correct Ans - Economical Order Quantity

Which of the following are elements of a distribution system? Correct Ans
- patient charging, storage, and receiving

The most efficient distribution system is the Correct Ans - Total
exchange cart

The computerized locked supply system is a variation of the Correct Ans -
Par Level

Which distribution system requires the using department to be responsible
for maintaing inventory levels? Correct Ans - Demand

An ongoing inventory management system is known as a Correct Ans -
Perpetual Inventory System

In which distribution system are requisition filled and delivered as they are
received? Correct Ans - Demand

Inventory performance" (a rough indication of the effectiveness of the
inventory system) is evaluated by Correct Ans - percent of back order,
availability of critical items, and inventory turnover

,Which of the following is the MOST efficient and cost-effective method for
inventory control in the SPD ordering and data generation? Correct Ans -
Computerized Inventory Control System

The temperature in the sterile storage are should not exceed Correct Ans
- 75º F (24º C)

Shelving in the sterile storage area must be kept how far from the outside
walls? Correct Ans - 2 inches

The thickness of a sterility maintenance cover should be Correct Ans -
2 to 3 mils

The minimum number of air exchanges per hour in the sterile storage area
should be Correct Ans - 4

The relative humidity in the sterile storage area should not exceed Correct
Ans - 70%

Approximately how much space should be left between and item and the
inside edges of a peel pouch? Correct Ans - 1 inch

What is the maximum size and weight of a textile pack? Correct Ans -
12" x 12 x 20" 12 pounds

Which of the following packaging materials is acceptable for low-temperature
gas plasma and ozone sterilization Correct Ans - Tyvek

Basin sets should be prepared for sterilization with Correct Ans - an
absorbent material between each basin

Tamper-evident seal on rigid containers include Correct Ans - latches
and locks

What is the acceptable temperature range for the prep and packaging area?
Correct Ans - 68º F-73º F (20º-23º C)

The preparation and packaging area must have an air exchange rate of
Correct Ans - 10 per hr

,If a non-perforated tray is used for the sterilization of instrument sets, it must
be Correct Ans - Tilted on the sterilizer rack

The ideal humidity level in the preparation and packing area should be
Correct Ans - 50%

For sterilization, rigid containers must be placed Correct Ans - flat

Which of the following packaging materials is acceptable for steam
sterilization? Correct Ans - Non-woven wraps

Re-passivation of an instrument can be achieved by Correct Ans -
routine lubrication

What is the most difficult part of a hemostat to clean? Correct Ans -
box lock

Which of the following lubricants is recommended for use on surgical
instruments? Correct Ans - petroleum-based

What type of finish on a surgical instrument is used for laser procedures?
Correct Ans - ebonized

What is the final stage in manufacturing a surgical grade instrument?
Correct Ans - passivation

Which series of stainless steel is used is to manufacture malleable retractors?
Correct Ans - 300

The weakest part of the instrument is/are the Correct Ans - box locks

The layer on a surgical instrument created by the passivation process is called
Correct Ans - chromium oxide

A crack in a hand-held instrument would most likely be found in the
Correct Ans - box lock

, The appropriate sterilization cycle for an autoclave load containing a basin
set, a minor surgery set, and twelve towel packs is Correct Ans - 270 F
(132 C ) for 4 minutes, 30 psi and 20 minutes drying time

Cotton gauze cannot be used in which sterilization process? Correct Ans -
low temperature gas plasma

Which agencies are involved with the regulation and control of ethylene oxide
when it's used as a sterilant? Correct Ans - environmental protection
agency, food and drug administration, and centers for disease control and
prevention

Which of the following sterilization methods would be appropriate for
petrolatum gauze dressings? Correct Ans - dry heat

To eliminate the bioburden on metal surgical instrument within the SPD, the
MOST cost-efficient sterilization method is generally considered to be
Correct Ans - steam under pressure

There are two pressure gauges on a steam sterilizer. These gauges measure
the pressure in the Correct Ans - jacket and chamber

What is the maximum number of 100% EO cartridges permitted by OSHA to
be stored outside of a flammable goods cabinet? Correct Ans - 10

At the end of the sterilization cycle, the EO cartridges and the gloves used to
handle them should be Correct Ans - Aerated

Item to be sterilized in low-temperature gas plasma must be packaged in:
Correct Ans - Non-cellulose-based material

Generally, the EO biological test pack is placed in the Correct Ans -
Geometric center of the load

Which of the following devices are compatible with ozone sterilization?
Correct Ans - stainless steel instruments

Routine BI testing for steam sterilizers should be performed at least
Correct Ans - weekly

Información del documento

Subido en
8 de julio de 2023
Número de páginas
37
Escrito en
2022/2023
Tipo
Examen
Contiene
Preguntas y respuestas
$16.49

¿Documento equivocado? Cámbialo gratis Dentro de los 14 días posteriores a la compra y antes de descargarlo, puedes elegir otro documento. Puedes gastar el importe de nuevo.
Escrito por estudiantes que aprobaron
Inmediatamente disponible después del pago
Leer en línea o como PDF

Seller avatar
Los indicadores de reputación están sujetos a la cantidad de artículos vendidos por una tarifa y las reseñas que ha recibido por esos documentos. Hay tres niveles: Bronce, Plata y Oro. Cuanto mayor reputación, más podrás confiar en la calidad del trabajo del vendedor.
StudyHall
3.8
(231)
Vendido
1358
Seguidores
825
Artículos
17365
Última venta
4 días hace




Por qué los estudiantes eligen Stuvia

Creado por compañeros estudiantes, verificado por reseñas

Calidad en la que puedes confiar: escrito por estudiantes que aprobaron y evaluado por otros que han usado estos resúmenes.

¿No estás satisfecho? Elige otro documento

¡No te preocupes! Puedes elegir directamente otro documento que se ajuste mejor a lo que buscas.

Paga como quieras, empieza a estudiar al instante

Sin suscripción, sin compromisos. Paga como estés acostumbrado con tarjeta de crédito y descarga tu documento PDF inmediatamente.

Student with book image

“Comprado, descargado y aprobado. Así de fácil puede ser.”

Alisha Student

Preguntas frecuentes