Minimal sedation 3.3.66.4 page 22 - ✔✔ correct answer A drug induced state during which patients
respond normally to verbal commands although cognitive functions and coordination may be
impaired, ventilatory and cardiovascular functions are unaffected.
Category two space 3.3.136.2 page 24 - ✔✔ correct answer Space in which failure of equipment or
system is likely to cause MINOR injury to patient, staff, or a visitor.
Category three space 3.3.136.3 page 24 - ✔✔ correct answer Space in which the failure of
equipment or a system is NOT LIKELY to cause injury to patient, staff, or visitor but can cause
discomfort.
Risk category 2 3.3.158.2 page 25 - ✔✔ correct answer Activities, systems or equipment who is a
failure is likely to cause MINOR injury to patient, staff, or visitors
Risk category three3.3.1 58.3 page 26 - ✔✔ correct answer Activities, systems, or equipment whose
failure is NOT LIKELY to cause injury to patient, staff, or visitors but can cause discomfort.
Primary supply 3.3.171.2 page 26 - ✔✔ correct answer The portion of the central supply system that
is default system for the piping system.
Reserve supply 3.3.171.3 page 26 - ✔✔ correct answer Where provided, the portion of the central
supply system that will supply the piping system win the primary and secondary supplies are
exhausted or are not operative.
Waste anesthetic gas disposal 3.3.181 page 26 - ✔✔ correct answer The process of capturing in
carrying our way gas is vented from the patient's breathing circuit during normal operation of gas
anesthetic or analgesia equipment.
Category 2. 4.1.2 page 27 - ✔✔ correct answer Activities, systems, or equipment who is failure is
likely to cause MINOR injury of patience, staff or visitors Shelby design to meet category choose
requirement, as details in this code.
Authority having jurisdiction 5.1.1.4 page 28 - ✔✔ correct answer And existing system that is not in
strict compliance with the provisions of this coach I'll be permitted to be continued and use as long
as the ______________________has determined that such use does not constitute just think hazard
to life
, 2018 NFPA 99 Medical Gas 6020
2 inches, 360°, 5.1.3.1.3 page 28 - ✔✔ correct answer Liquid containers shall have additional product
identification visible from all directions with a minimum of ________high letters such as a
___________wraparound tape for medical liquid containers.
125°, 20°, 5.1.3.2.12 and 13 page 29 - ✔✔ correct answer Cylinders in use and in storage shall be
prevented from reaching temperatures in excess of ________and shall never be lower than _______.
Manifold for gas cylinders, manifold for cryo-genic liquid containers, in building emergency reserves,
instrument air stand by headers, individual components on the actual inside of concentrator sources
5.1.3.3.1.2 page 29 - ✔✔ correct answer What are the following systems shall be permitted to be
located together in the same INDOOR enclosure?
Medical errors central supply compressor supply stores, medical surgical vacuum central supply
stores, waste anesthetic gas disposal supply sources..5.1.3.3.1.3 page 29 - ✔✔ correct answer What
are the following CENTRAL SUPPLY systems shall be permitted to be located together in the same
room?
200 ft.² 5.1.3.3.2 page 29 - ✔✔ correct answer During the design and construction I have a central
supply system and the storage of positive pressure Is located outdoors and is greater than
_________, they shall be provided with a minimum of two entry/exit.
Motor driven machinery 5.1.3.3.4.2 page 30 - ✔✔ correct answer Cylinders, whether full or empty,
shall not be stored in enclosure containing ___________, with the exception of cylinders intended
for instrument air reserve headers, which shall be permitted to be placed in the same location
containing an instrument air compressor when it has motor driven machinery located within the
room. All the cylinders intended for the instrument air reserve header come shot be permitted to be
stored in enclosure containing instrument air compressors
350 psi 5.1.3.5.4 page 31 - ✔✔ correct answer In those central supply systems, portions of the
system intended to handle oxygen at gauge pressure is greater than ________, enter connecting
hoses shall contain no polymeric materials
Brass, bronze, or stainless steel construction 5.1.3.5.6.1 page 31 - ✔✔ correct answer A pressure
relief valve shall have what construction?
A relief valve will be located between each final line regulator and source valve, they sure have a
setting a 50% above normal system operating pressure. 5.1.3.5.6.3 page 31 - ✔✔ correct answer
Central supply system for positive pressure relief valves, include one or more relief valve, all meeting
the following requirements