QUESTIONS AND VERIFIED CORRECT ANSWERS
An alarm indicating a partial exposure (when the trigger was
not depressed through the entire timer setting, either on
purpose or accidentally) is indicated by a series of beeps and
the ____________ message on the screen until a button is
pressed. - ---Answers----INCOMPETE / RED
Do not ____ the NOMAD Pro 2 until the patient and film/ sensor
are positioned and ready to avoid unintended firing of the
NOMAD Pro 2. - ---Answers----ENABLE / READY
When there is not enough battery charge left to take an
exposure, the charge indicator bars will fade away and the ____
alarm message appears on the screen, and the NOMAD Pro 2
will not allow an exposure to begin. - ---Answers----
RECHARGE HANDSET
If left unused NOMAD Pro 2 will automatically shut down after
approximately
______________ minute(s). The NOMAD Pro 2 can also be turned
off by pressing the POWER button. - ---Answers----3 minutes
, Some battery ____ may be lost during extended inactivity
(leading to fewer exposures between each Handset charging
session). - ---Answers----charge
In order to maintain the optimal operator safety zone the
backscatter shield should be:
a) held so it is parallel to the oeprator
b) slid back on the cone toward the operator
c) held so NOMAD Pro 2 is positioned close to the patient - ---
Answers----a) Held so it is parallel to the operator
c) Held so NOMAD Pro 2 is positioned close to the patient.
The duty cycle (or rest period b/w exposures) to protect your
NOMAD Pro2 is as follows:
a) 6 sec for an exposure 0.09 sec long
b) 12 sec for an exposure 0.20 sec long
c) 24 sec for an exposure 0.40 sec long
d) 45 sec for an exposure 1.00 sec long - ---Answers----a) 6
sec for an exposure 0.09 sec long
b) 12 sec for an exposure 0.20 sec long
c) 24 sec for an exposure 0.40 sec long
The Technique Factor for making exposures:
a) Is represented in a chart that is for reference only in
establishing a customized exposure technique in your office.