Ohm's law equation - CORRECT ANSWER-V=IR
EKG monitored patient should be __________, monitor chassis should be
___________ (nonconductive). Why - CORRECT ANSWER-ungrounded,
grounded. Grounded patient would be able to conduct leakage currents. we
dont want this
lb to kg - CORRECT ANSWER-2.2 lb = 1 kg
F to C - CORRECT ANSWER-C=5/9(F-32)
C to F - CORRECT ANSWER-F=9/5C+32
3 things that change automaticity of automatic cells - CORRECT ANSWER-1.
slope of phase 4
2. Change of threshold potential
3. Change of resting membrane potential
a systole is represented by what portion of the ECG - CORRECT ANSWER-QT
interval
,What is main ion transfer during systole - CORRECT ANSWER-K+ seeps out,
Ca2+ enters
At rest, what is intracellular/extracellular environment - CORRECT ANSWER-
inside negative, outside positive
What phase of AP is resting membrane potential - CORRECT ANSWER-phase
4
Nodal depolarization occurs via what channel - CORRECT ANSWER-Ca2+ slow
channel
Cardiac muscle cell depol occurs via what channel - CORRECT ANSWER-Na+
fast channel
During which phases is cell refractory - CORRECT ANSWER-1,2,3
TDP caused by what mechanism usually - CORRECT ANSWER-EAD's (triggered
activity)
Variance - CORRECT ANSWER-Take each number in sample, subtract each
number from mean, square each difference, add all, divide sum by number in
sample -1
Standard deviation - CORRECT ANSWER-square root of variance
P value - CORRECT ANSWER-Probability of certainty / smaller p value means
the more likely the result could not occur by chance
, survival curve showing % of patients surviving treatment over time - CORRECT
ANSWER-Kaplan - Meier curves
when measuring EGM's, earliest atrial activation usually from _____ and
earliest ventricular usually from ______ - CORRECT ANSWER-P wave / QRS
Morady maneuver is used to differentiate - CORRECT ANSWER-differentiate
AT from AVNRT/AVRT
Most common arrhythmia - CORRECT ANSWER-AF
Most common SVT - CORRECT ANSWER-AVNRT
What type of single use devices can be resterilized and why? - CORRECT
ANSWER-Diagnostic EP electrodes because lumen and technically not a
catheter because no lumen
Fr to mm - CORRECT ANSWER-1 Fr = .33 mm
Tip inner diameter standard diagnostic cardiac catheters and why - CORRECT
ANSWER-.038 inch / because it is supposed to be used with .035-.038 inch
guide wires
What is 2 rules for CMC's like spiral or lasso - CORRECT ANSWER-1. Only
approved for use in LA due to chordae tendinea
2. Only rotate shaft clockwise