ASSE 6005 Certification EXAM QUESTIONS AND ANSWERS
Likely to cause major inury or death of patient or caregiver - CORRECT ANSWER✅✅Category 1
Likely to cause minor injury to patient or caregiver - CORRECT ANSWER✅✅Category 2
Not likely to cause injury, but may cause patient discomfort - CORRECT ANSWER✅✅Category 3
No impact on patient care - CORRECT ANSWER✅✅Category 4
Within an area of use, provides continuous audible and visible surveillance of Level 1 and 2 systems -
CORRECT ANSWER✅✅Area Alarm System
Provides continuous audible and visible surveillance of system source equipment at equipment site -
CORRECT ANSWER✅✅Local Alarm System
Monitors operation and condition of the source of supply, the reserve source (if any), and the pressure
in the main lines of each system - CORRECT ANSWER✅✅Master Alarm System
Special care units, intensive care units, coronary care units, angiography laboratories, cardiac
catheterization laboratories, delivery rooms, emergency departments, and similar areas in which
patients are intended to be subjected to invasive procedures and connected to line-operated, patient
care related electrical devices - CORRECT ANSWER✅✅Critical Care Areas
Medical Gas Systems Generalist - CORRECT ANSWER✅✅ASSE 6005
Medical Gas Systems Installer - CORRECT ANSWER✅✅ASSE 6010
Bulk Medical Gas Systems Installer - CORRECT ANSWER✅✅ASSE 6015
, Medical Gas Systems Inspector - CORRECT ANSWER✅✅ASSE 6020
Medical Gas Systems Verifier - CORRECT ANSWER✅✅ASSE 6030
Bulk Medical Gas Systems Verifier - CORRECT ANSWER✅✅ASSE 6035
Medical Gas Systems Maintenance Personnel - CORRECT ANSWER✅✅ASSE 6040
Medical Gas Systems Instructor - CORRECT ANSWER✅✅ASSE 6050
Bulk Medical Gas Systems Instructor - CORRECT ANSWER✅✅ASSE 6055
High pressure cylinder delivery pressure - CORRECT ANSWER✅✅50-55 psi
Shutoff valve at immediate connection of each source to piped distribution system - CORRECT
ANSWER✅✅Source Valve
A fluid with a boiling point lower than -130 deg F at an absolute pressure of 14.7 psi - CORRECT
ANSWER✅✅Cryogenic Fluid
Shutoff valve in main supply line inside of the building - CORRECT ANSWER✅✅Main Line Valve
Instrument air design pressure - CORRECT ANSWER✅✅200 psi
Shutoff valve in each riser adjacent to main line - CORRECT ANSWER✅✅Riser Valve
Shutoff valve in each lateral branch from main or riser - CORRECT ANSWER✅✅Service Valve
Shutoff valves for station outlets and inlets - CORRECT ANSWER✅✅Zone Valve
Likely to cause major inury or death of patient or caregiver - CORRECT ANSWER✅✅Category 1
Likely to cause minor injury to patient or caregiver - CORRECT ANSWER✅✅Category 2
Not likely to cause injury, but may cause patient discomfort - CORRECT ANSWER✅✅Category 3
No impact on patient care - CORRECT ANSWER✅✅Category 4
Within an area of use, provides continuous audible and visible surveillance of Level 1 and 2 systems -
CORRECT ANSWER✅✅Area Alarm System
Provides continuous audible and visible surveillance of system source equipment at equipment site -
CORRECT ANSWER✅✅Local Alarm System
Monitors operation and condition of the source of supply, the reserve source (if any), and the pressure
in the main lines of each system - CORRECT ANSWER✅✅Master Alarm System
Special care units, intensive care units, coronary care units, angiography laboratories, cardiac
catheterization laboratories, delivery rooms, emergency departments, and similar areas in which
patients are intended to be subjected to invasive procedures and connected to line-operated, patient
care related electrical devices - CORRECT ANSWER✅✅Critical Care Areas
Medical Gas Systems Generalist - CORRECT ANSWER✅✅ASSE 6005
Medical Gas Systems Installer - CORRECT ANSWER✅✅ASSE 6010
Bulk Medical Gas Systems Installer - CORRECT ANSWER✅✅ASSE 6015
, Medical Gas Systems Inspector - CORRECT ANSWER✅✅ASSE 6020
Medical Gas Systems Verifier - CORRECT ANSWER✅✅ASSE 6030
Bulk Medical Gas Systems Verifier - CORRECT ANSWER✅✅ASSE 6035
Medical Gas Systems Maintenance Personnel - CORRECT ANSWER✅✅ASSE 6040
Medical Gas Systems Instructor - CORRECT ANSWER✅✅ASSE 6050
Bulk Medical Gas Systems Instructor - CORRECT ANSWER✅✅ASSE 6055
High pressure cylinder delivery pressure - CORRECT ANSWER✅✅50-55 psi
Shutoff valve at immediate connection of each source to piped distribution system - CORRECT
ANSWER✅✅Source Valve
A fluid with a boiling point lower than -130 deg F at an absolute pressure of 14.7 psi - CORRECT
ANSWER✅✅Cryogenic Fluid
Shutoff valve in main supply line inside of the building - CORRECT ANSWER✅✅Main Line Valve
Instrument air design pressure - CORRECT ANSWER✅✅200 psi
Shutoff valve in each riser adjacent to main line - CORRECT ANSWER✅✅Riser Valve
Shutoff valve in each lateral branch from main or riser - CORRECT ANSWER✅✅Service Valve
Shutoff valves for station outlets and inlets - CORRECT ANSWER✅✅Zone Valve