Identify the components of the low-pressure system in the
anesthesia machine (Select 2):
- flowmeter tubes
- oxygen flush valve
- common gas outlet
- cylinder pressure regulator - CORRECT ANSWER >>>>>-
common gas outlet
- flowmeter tubes
The anesthesia machine can be divided into the high,
intermediate, and low pressure systems.
High pressure = ________ pressure
Begins at the ___________.
Ends at the _______________________.
Components:
-
-
,-
- - CORRECT ANSWER >>>>>cylinder pressure
cylinder
cylinder pressure regulator
- hanger yoke
- yoke block with check valves
- cylinder pressure gauge
- cylinder pressure regulator
Intermediate-pressure system:
Pipeline = _____ psi
Tank = _____ psi
Begins at the ____________.
Ends at the _______________________.
Components:
-
-
-
,-
-
-
- - CORRECT ANSWER >>>>>50 psi
45 psi
pipeline
flow-control valves
- pipeline inlets
- pipeline pressure gauge
- oxygen supply failure device/alarm
- second stage oxygen regulator
- O2 flush valve
- ventilator power inlet
- flowmeter control valves
Low-pressure system:
Pressure is __________________________________, but
variable.
Begins at the ___________________.
, Ends at the _____________________.
Components:
-
-
-
- - CORRECT ANSWER >>>>>slightly above atmospheric
flowmeter tubes
common gas outlet
flowmeter tubes (Thrope tubes)
Vaporizers
One-way Check valve
Common gas outlet
Low-pressure Leak Test:
(Negative Pressure Test)
The low-pressure leak test (negative pressure test) assesses the
integrity of the low-pressure system from the flowmeter valves
to the common gas outlet.
with Answers
Identify the components of the low-pressure system in the
anesthesia machine (Select 2):
- flowmeter tubes
- oxygen flush valve
- common gas outlet
- cylinder pressure regulator - CORRECT ANSWER >>>>>-
common gas outlet
- flowmeter tubes
The anesthesia machine can be divided into the high,
intermediate, and low pressure systems.
High pressure = ________ pressure
Begins at the ___________.
Ends at the _______________________.
Components:
-
-
,-
- - CORRECT ANSWER >>>>>cylinder pressure
cylinder
cylinder pressure regulator
- hanger yoke
- yoke block with check valves
- cylinder pressure gauge
- cylinder pressure regulator
Intermediate-pressure system:
Pipeline = _____ psi
Tank = _____ psi
Begins at the ____________.
Ends at the _______________________.
Components:
-
-
-
,-
-
-
- - CORRECT ANSWER >>>>>50 psi
45 psi
pipeline
flow-control valves
- pipeline inlets
- pipeline pressure gauge
- oxygen supply failure device/alarm
- second stage oxygen regulator
- O2 flush valve
- ventilator power inlet
- flowmeter control valves
Low-pressure system:
Pressure is __________________________________, but
variable.
Begins at the ___________________.
, Ends at the _____________________.
Components:
-
-
-
- - CORRECT ANSWER >>>>>slightly above atmospheric
flowmeter tubes
common gas outlet
flowmeter tubes (Thrope tubes)
Vaporizers
One-way Check valve
Common gas outlet
Low-pressure Leak Test:
(Negative Pressure Test)
The low-pressure leak test (negative pressure test) assesses the
integrity of the low-pressure system from the flowmeter valves
to the common gas outlet.