SEE exam-Anesthesia Machine
Questions with 100% Correct Answers
High pressure system Correct Answer: The cylinder pressure regululator is a part
of what system?
Intermediate-pressure Correct Answer: The oxygen flush valve is part of what
system?
Low-pressure Correct Answer: The flowmeter tubes and the common gas outlet
are part of what system?
Begins-cylinder
Ends-Cylinder regulators Correct Answer: Where does the high-pressure system
begin and end?
Begins-Pipeline
Ends-Flowmeter valves Correct Answer: Where does the intermediate-pressure
system begin and end?
There's a leak in the system Correct Answer: If the bulb inflates while doing a
low-pressure test, what does that indicate?
Integrity of low-pressure circuit from flowmeter valves to common gas outlet
Correct Answer: What does the low-pressure leak test assess?
Close APL valve, turn pressure to 30 cmH20, and pressure should stay at 30
cmH20 Correct Answer: How do you perform a high-pressure leak test?
1. O2 pressure alarm
Questions with 100% Correct Answers
High pressure system Correct Answer: The cylinder pressure regululator is a part
of what system?
Intermediate-pressure Correct Answer: The oxygen flush valve is part of what
system?
Low-pressure Correct Answer: The flowmeter tubes and the common gas outlet
are part of what system?
Begins-cylinder
Ends-Cylinder regulators Correct Answer: Where does the high-pressure system
begin and end?
Begins-Pipeline
Ends-Flowmeter valves Correct Answer: Where does the intermediate-pressure
system begin and end?
There's a leak in the system Correct Answer: If the bulb inflates while doing a
low-pressure test, what does that indicate?
Integrity of low-pressure circuit from flowmeter valves to common gas outlet
Correct Answer: What does the low-pressure leak test assess?
Close APL valve, turn pressure to 30 cmH20, and pressure should stay at 30
cmH20 Correct Answer: How do you perform a high-pressure leak test?
1. O2 pressure alarm