100% GRADED |100 REAL & COMPLETE QUESTIONS AND ANSWERS
| 2026 VERSION , GET IT CORRECT!!!
What pulse sequence will produce the most patient heating?
A. Inverse recovery sequence
B. Single shot fast spin echo
C. Conventional spin echo
D. Gradient Echo - (ANSWER)B. Single shot fast spin echo
During acquisition, it is noticed that the T wave of the cardiac cycle is elevated. What is this
artifact called?
A. Magnetohydrodynamic Effect
B. Excitation
C. Teratogenesis
D. Magnetophosphenes - (ANSWER)A. Magnetohyrdodynamic Effect
When can heating occur in a wire?
,A. Only in ferrous object
B. When the transmitted wavelength is half of the B0 field strength or is a multiple of the length
of that
C. When objects get smaller
D. In small round metal objects - (ANSWER)B. When the transmitted wavelength is half of the
B0 field strength or is a multiple of the length of that
What unit of measure is used for SAR?
A. A/1b
B. dB/dx
C. dB/dt
D. W/kg - (ANSWER)D. W/kg
The IEC 60601-2-33 whole body normal operating mode is ________________ the energy
requirements limited in the first level controlled operating mode.
A. Triple
,B. Double
C. Half
D. Not comparable to - (ANSWER)C. Half
Which of the following are recommended by the ACR guidance in MR Safety 2013 in proper
scanning of a patient?
A. Metal Detector
B. Ferromagnetic detection system
C. Written screening sheet
D. Patient Gown - (ANSWER)B. Ferromagnetic detection system
C. Written screening sheet
D. Patient Gown
Using a headset and ear plugs will reduce the noise by how much?
A. By the NRR of headset
B. By the NRR of the headset and NRR of earplugs
C. 50 dB
, D. By the NRR of the headset and the NRR of earplug divided by 2 - (ANSWER)B. By the NRR
of the headset and NRR of earplugs
What is the FDA SAR limit for the whole body in first level controlled operating mode?
A. 7 W/kg for 6 min
B. 4 W/kg for 6min
C. 3 W/kg for 6 min
D. 8 W/kg for 6 min - (ANSWER)B. 4 W/kg for 6 min
Which of the following is true of Zone IV? Select all that apply.
A. Should be noted with signage
B. No unsafe implants may cross the 5-guass unless implant is researched and analyzed for risk
C. Contain metal detectors for implant identification
D. This area should be monitored by a level 2 MR personnel - (ANSWER)A. Should be noted
with signage
B. No unsafe implants may cross the 5-guass unless implant is researched and analyzed for risk
D. This area should be monitored by a level 2 MR personnel