IBHRE CCDS EXAM NEWEST 2025-2026
UPDATED EXAM ALL QUESTIONS &
ANSWERS JUST RELEASED
IBHRE CCDS ACTUAL EXAM
A 47 yr/old male who works for Verizon experiences dyspnea while servicing
cell phones in the office. Stored EGM's of his 2 yr/old bipolar DDDR device
reveal some pauses. Which of the following is the most likely cause?
a) ventricular oversensing of the cell phone
b) atrial oversensing of the cell phone
c) ventricular oversensing of myopotential
d) ventricular insulation break
D. Ventricular Insulation Break
Which of the following parameters would indicate a good time to increase the
follow-up frequency from every 6 months to every 2 months in a DDDR
system?
a) battery voltage 2.2 V
b) battery voltage 2.7 V
c) cell impedance 325 ohms
d) cell impedance 3.7 Kohms
D. Cell Impedance 3.7 Kohms
Which of the following would be most accurate when contrasting the
subclavian approach to the cephalic approach of transvenous lead
introduction?
a) hemothorax-high, pneumothorax-high, multiple leads-low
b) crush-high, thoracic duct injury-high, pneumothorax-low
c) brachial plexus injury-high, crush-low, multiple leads high
d) lung laceration-high, crush-high, arterial puncture-high
,D. Lung Laceration High, Crush High, Arterial Puncture High
Which of the following is the most common reason for a pacemaker
classifying a false VT/VF episode?
a) atrial undersensing
b) T-wave over-sensing
c) R-wave double counting
d) electromagnetic interference
A. Atrial Undersensing
Ability to mode switch in the presence of fine atrial fibrillation may be
enhanced by each of the following EXCEPT:
a) increase atrial sensitivity
b) decrease mode switch rate
c) decrease PVAB
d) decrease PVARP
D. Decrease PVARP
,A 48 yr/old pacemaker dependent woman golfer is scheduled for a shoulder
repair where electrocautery will be utilized. The battery measures 2.56 V and
ERI is 2.5
V. Which of the following is the best course of action?
a) program the device to asynchronous mode
b) place the grounding plate as far away as possible limit electrocautery to
short bursts
c) replace the device before surgery
D. Replace the Device Before Surgery
d)
Which of the following explains the almost 50 year history of RV apical
pacing?
a) improved lead stability
b) reduced intraventricular conduction times
c) reduced ventricular conduction times
d) improved systolic fiber shortening
A. Improved Lead Stability
, In a pacing system that fails to pace bipolar but paces normally in the unipolar
configuration, the most likely explanation would be:
a) loose anodal set-screw
b) loose cathodal set-screw
c) outer insulation break
d) inner insulation break
A. Loose Anodal Set Screw
Approximately, what % of sinus node disease patients develop AV block?
a) 1-1.5%/yr
b) 3-5%/yr
c) 5-10%/yr
d) 10-15%/yr
A. 1-1.5% per year
A 25 year old idiopathic cardiomyopathy patient, with an EF of 15%, is
implanted with an ICD. At the time of implant, there is a concern that he may
be too unstable to tolerate DFT testing. An advantage of performing the DFT
testing would be:
a) Check integrity of the system
b) Check lead connections
c) Confirm adequate VF sensing
d) All of the above
D. All the above
UPDATED EXAM ALL QUESTIONS &
ANSWERS JUST RELEASED
IBHRE CCDS ACTUAL EXAM
A 47 yr/old male who works for Verizon experiences dyspnea while servicing
cell phones in the office. Stored EGM's of his 2 yr/old bipolar DDDR device
reveal some pauses. Which of the following is the most likely cause?
a) ventricular oversensing of the cell phone
b) atrial oversensing of the cell phone
c) ventricular oversensing of myopotential
d) ventricular insulation break
D. Ventricular Insulation Break
Which of the following parameters would indicate a good time to increase the
follow-up frequency from every 6 months to every 2 months in a DDDR
system?
a) battery voltage 2.2 V
b) battery voltage 2.7 V
c) cell impedance 325 ohms
d) cell impedance 3.7 Kohms
D. Cell Impedance 3.7 Kohms
Which of the following would be most accurate when contrasting the
subclavian approach to the cephalic approach of transvenous lead
introduction?
a) hemothorax-high, pneumothorax-high, multiple leads-low
b) crush-high, thoracic duct injury-high, pneumothorax-low
c) brachial plexus injury-high, crush-low, multiple leads high
d) lung laceration-high, crush-high, arterial puncture-high
,D. Lung Laceration High, Crush High, Arterial Puncture High
Which of the following is the most common reason for a pacemaker
classifying a false VT/VF episode?
a) atrial undersensing
b) T-wave over-sensing
c) R-wave double counting
d) electromagnetic interference
A. Atrial Undersensing
Ability to mode switch in the presence of fine atrial fibrillation may be
enhanced by each of the following EXCEPT:
a) increase atrial sensitivity
b) decrease mode switch rate
c) decrease PVAB
d) decrease PVARP
D. Decrease PVARP
,A 48 yr/old pacemaker dependent woman golfer is scheduled for a shoulder
repair where electrocautery will be utilized. The battery measures 2.56 V and
ERI is 2.5
V. Which of the following is the best course of action?
a) program the device to asynchronous mode
b) place the grounding plate as far away as possible limit electrocautery to
short bursts
c) replace the device before surgery
D. Replace the Device Before Surgery
d)
Which of the following explains the almost 50 year history of RV apical
pacing?
a) improved lead stability
b) reduced intraventricular conduction times
c) reduced ventricular conduction times
d) improved systolic fiber shortening
A. Improved Lead Stability
, In a pacing system that fails to pace bipolar but paces normally in the unipolar
configuration, the most likely explanation would be:
a) loose anodal set-screw
b) loose cathodal set-screw
c) outer insulation break
d) inner insulation break
A. Loose Anodal Set Screw
Approximately, what % of sinus node disease patients develop AV block?
a) 1-1.5%/yr
b) 3-5%/yr
c) 5-10%/yr
d) 10-15%/yr
A. 1-1.5% per year
A 25 year old idiopathic cardiomyopathy patient, with an EF of 15%, is
implanted with an ICD. At the time of implant, there is a concern that he may
be too unstable to tolerate DFT testing. An advantage of performing the DFT
testing would be:
a) Check integrity of the system
b) Check lead connections
c) Confirm adequate VF sensing
d) All of the above
D. All the above